Workshop on certificate of dew
Download
Report
Transcript Workshop on certificate of dew
คำนวณหำควำมชื ้นสัมพัทธ์ที่ถกู ต้ องใน Calibration Certificate จำกหน่วย
dew-point temperature
?
% RH
?
?
e ' w ( t s ) f ( t s , Ps )
e ' w ( t c ) f ( t c , Pc )
Pa
Pa
100
?
where
and e’w(ts) and e’w(tc) are the saturation water vapor pressure over a plane surface of the
pure phase of liquid water or solid ice at the saturator temperature ts and test chamber
temperature tc
The temperature and pressure have units of Kelvin and Pascal
f ( Ps,ts ) is the enhancement factor of water vapor-air mixture at saturator pressure Ps
and temperature ts
f ( Pc ,tc) is the enhancement factor of water vapor-air mixture at test chamber pressure
Pc and temperature tc
P is Pressure on temperature chamber (Pc) and temperature on saturation (Ps)
เรารู้อะไรบ้าง ?
Dew-point/Frost-point temperature ( oC)
ได้จาก ?
2. Ambient temperture ( oC)
1.
3.
ได้จาก ?
สูตรการคานวณ ew และ f
Bob Hardy, Formulation for vapor pressure…,
Institute of measurement control, A guide to the measurement of humidity
ฯลฯ
Dew-point/ Frostpoint temperature
Dew-point
hygrometer
Pout = 101325 Pa
td = Average of calibration
dew-point temperature
reading
ta = Average of calibration
ambient temperature
reading
ta &td & Pa
ค่าเฉลี่ยจากการบันทึกผล
การวัด
td
ed '
e w ( t d ) f ( t d , Pa )
คานวนหา ew และ f ของ td
ta
ea '
e w ( t a ) f ( t a , Pa )
e '
% RH d
e '
a
P
a 100
P
a
คานวนหา ew และ f ของ ta
Actual water
vapor pressure
e' e f
ตัวแปรที่ไม่ทราบค่า (1)
By Wexler’s equation and Sonntag’s equation;
6
i 2
e w exp ai (T 273.15) a ln(T 273.15)
7
i
0
By Magnus’s equation
a t
e exp lna 1
a t
w
0
2
Wexler
Range: above water
Sonntag
Hardy
Wagner and Pruss
ITS-90
Magnus
a0
a1
a2
a3
a4
a5
a6
a7
-2.9912729·103
-6.0170128·103
1.887643854·101
-2.8354721·10-2
1.7838301·10-5
-8.4150417·10-10
4.4412543·10-13
2.858487
0
-6.0969385·103
2.12409642·101
-2.711193·10-2
1.673952·10-5
0
0
2.433502
-2.8365744·103
-6.028076559·103
1.954263612·101
-2.737830188·10-2
1.6261698·10-5
7.0229056·10-10
-1.8680009·10-13
2.7150305
0
-7.85951783
1.84408259
-11.7866497
22.6807411
-15.9618719
1.80122502
0
611.2
17.62
243.12
ur(e)
for the range
0oC ≤ t ≤ 100oC:
ur(e) < 0.005%
as for Wexler
as for Wexler
for the range
-45 oC ≤ t ≤60oC:
ur(e) < 0.3%
a0
a1
a2
a3
a4
a5
a6
a7
ur(e)
Wexler
above ice Equation
Sonntag
Hardy
Wagner and Pruss
ITS-90
Magnus
0
-5.6745359·103
-5.6745359
-9.677843·10-3
6.22157·10-7
2.0747825·10-9
-9.484024·10-13
4.1635019
0
-6.0245282·103
2.932707·101
1.0613868·10-2
-1.3198825·10-5
0
0
-4.9382577·10-1
0
-5.8666426·103
2.232870244·101
1.39387003·10-2
-3.4262402·10-5
2.7040955·10-8
0
6.7063522·10-1
0
-13.928169
34.7078238
611.2
22.46
272.62
for the range
-100oC ≤ t ≤
0.01oC:
ur(e) < (0.010.005t)% of value
as for Wexler
as for Wexler
for the range
-65 oC ≤ t ≤0.01oC:
ur(e) < 0.5%
Actual water
vapor pressure
ตัวแปรที่ไม่ทราบค่า (2)
e' e f
P
ed
Water pressure enhancement factor f exp α 1 β 1
P
ed
4
4
i 1
Ai t
( i 1)
i 1
Bi t
( i 1)
ไอนำ้ เป็ นแก๊สอุดมคติหรือไม่ ?
Sensitivity Coefficient (Ci)
สามารถหาได้จาก
For Dew-point Temperature =
dRH
dt d
for magnus
equation
dRH
dt d
for sonntag
equation
dRH
dt d
4283 . 774
RH
243 . 12 t d
2
a7
2
RH a 1T d a 3 2 a 4 T d
T
d
Sensitivity Coefficient (Ci)
for Wexler
equation
dRH
dt d
a7
3
2
2
3
RH 2 a 0 T d a 1T d a 3 2 a 4 T d 3 a 5T d 4 a 6 T d
T
d
for air temper ature sensitivit
y coefficien
t
dRH
dT a
for magnus
equation
dRH
dt a
for sonntag
equation
dRH
dt a
4283 . 774
- RH
2
243 . 12 t a
a7
2
- RH a 1T a a 3 2 a 4 T a
T
a
Sensitivity Coefficient (Ci)
for Wexler
equation
dRH
dt a
a7
3
2
2
3
RH 2 a 0 T a a 1T a a 3 2 a 4 T a 3 a 5 T a 4 a 6 T a
T
a