All Ions MS/MS
1
High Productivity Forensic
Screening with More
Confident Identification
Confidentiality Label
July 7, 2015
Topics
Screening vs.
Targeted Analysis
What is All Ions
MS/MS?
Drug Screening with
All Ions MS/MS
2
Confidentiality Label
July 7, 2015
Non-Target Screening vs. Targeted Quantitation
Targeted Quantitation
Non-Target Screening
LC/QQQ
LC/TOF or LC/Q-TOF
Benefits:
Analysis Options:
•
Most Sensitive Technique
•
MS only screening
•
Good Quantitative Results
•
MS/MS with accurate mass library
for confident identification
Drawbacks:
•
Limited number of analytes
•
Non-target compounds undetected
3
Benefits:
•
Detect, Quantify, and Identify
hundreds or thousands of
compounds in a single injection
•
Reanalyze old sample data any time
without reinjection upon discovery of
new compounds
Confidentiality Label
July 7, 2015
Non-Target Forensic Screening
The Value of the Agilent Solution
• Robust hardware
• Sample Preparation Products
• HPLC and UHPLC Columns
• Automated software workflows
• Comprehensive Accurate
Mass Libraries
• LC/MS Application Kits
• First-class service and
application support
4
Confidentiality Label
July 7, 2015
Agilent Forensic Toxicology PCDL
Database with 9,000+ Compounds
•
Designer Drugs
•
Hallucinogens
•
Stimulants
 Accurate Mass Correction
•
Opioids
 Molecular Structure
•
Benzodiazepines
•
Hypnotics
 Quality Controlled
•
Neuroleptics
•
Barbituates
•
Antidepressants
•
Antiepileptics
 CAS Number/PubChem Link
Library spectra for 3,000+ Compounds
5
Confidentiality Label
July 7, 2015
Topics
Screening vs.
Targeted Analysis
What is All Ions
MS/MS?
Drug Screening with
All Ions MS/MS
6
Confidentiality Label
July 7, 2015
All Ions MS/MS-What is it?
Step 1: Fragmentation without Isolation on TOF or Q-TOF
a) First Scan: Low fragmentation energy to analyze precursors
b) Second Scan: High fragmentation energy to analyze fragment ions
7
Low energy spectrum
Shows
Precursor
Ions
High energy spectrum
Shows
Fragment
Ions
Confidentiality Label
July 7, 2015
All Ions MS/MS-What is it?
Step 2: Software Extracts, Correlates, and Confirms
a) Find by Formula uses library to extract precursors from low energy MS
channel
b) All Ions MS/MS uses library to extract fragments from high energy
channel, gives coelution score
Software automatically
matches precursors with
fragment ions:
Coelution Plot with
Coelution Score
8
Confidentiality Label
July 7, 2015
All Ions MS/MS: What is the Value?
• More Confident Identification
• Simple Method Set Up
• High Productivity
Software Workflow
• Reanalyze Old Samples
Without Reinjection
to Find New Compounds
• Forensic Toxicology PCDL
and Application Kit with 9,000+
Forensic Compounds of Interest
9
Confidentiality Label
July 7, 2015
Topics
Screening vs.
Targeted Analysis
What is All Ions
MS/MS?
Drug Screening with
All Ions MS/MS
10
Confidentiality Label
July 7, 2015
Application of All Ions MS/MS
Untargeted Drug Screening in Postmortem Blood
Purpose of Application:
• Evaluate Efficiency of All Ions MS/MS
Workflow
• Unambiguously Identify Coeluting Analytes
at Different Concentrations
11
Confidentiality Label
July 7, 2015
Experimental
Untargeted Drug Screening in Postmortem Blood
• Preparation of post-mortem blood samples
• Liquid/liquid extraction with 90:10 acetonitrile:ethanol
• 250 mg blood + 500 µL extraction solvent
• 2 µL of extract injected
• Run time: 12 minutes
• Two Analyses:
• 30-compound test standard
• Blood sample spiked at 0.1 µg/g
12
Confidentiality Label
July 7, 2015
Evaluation of All Ions MS/MS Automated Workflow
Untargeted Drug Screening in Postmortem Blood
Step 1: Find by Formula Extracts Precursors using PCDL database
13
Confidentiality Label
July 7, 2015
Evaluation of All Ions MS/MS Automated Workflow
Untargeted Drug Screening in Postmortem Blood
Step 2: Extract Corresponding Fragment Ions
Fragments from PCDL
14
Confidentiality Label
July 7, 2015
Evaluation of All Ions MS/MS Automated Workflow
Untargeted Drug Screening in Postmortem Blood
Step 3: Precursor and Fragment Correlation/Coelution
Overlaid Precursor and
Fragment Ion
Chromatograms
Coelution Plot
15
Confidentiality Label
July 7, 2015
Differentiation and Identification of Isomers
Untargeted Drug Screening in Postmortem Blood
Alprazolam
Coeluting with
~50x
Concentration
Difference
Methadone
16
Confidentiality Label
July 7, 2015
Differentiation and Identification of Isomers
Untargeted Drug Screening in Postmortem Blood
Similar precursor
m/z values and
overlapping isotopic
clusters
17
Confidentiality Label
July 7, 2015
Compound Confirmation-Methadone
Untargeted Drug Screening in Postmortem Blood
Red outline overlays
theoretical isotopic pattern
and abundance with
acquired precursor data
for methadone
Methadone confirmed
with seven qualified
fragments
18
Confidentiality Label
July 7, 2015
Compound Confirmation-Alprazolam
Untargeted Drug Screening in Postmortem Blood
Red outline overlays
theoretical isotopic pattern
and abundance with
acquired precursor data
for alprazolam
Alprazolam confirmed
with two qualified
fragments, despite
low concentration
19
Confidentiality Label
July 7, 2015
High Productivity: Compound Details View
Compound List
Overlaid
Chromatograms
Compound ID Results
Low Energy (MS) Spectrum
High Energy Fragment Spectrum
Co-elution Plot
20
Confidentiality Label
July 7, 2015
Conclusions
• All Ions MS/MS Increases Confidence in ID
• Simple Method Setup
• Automated data processing
• Accurate Mass MS/MS Library Extraction
• Coelution Plot with Coelution Score
• Reanalyze data any time, without reinjection of samples
• Review data quickly, increasing productivity
21