Optical Quality of the Synchrony IOL
Download
Report
Transcript Optical Quality of the Synchrony IOL
The Synchrony IOL is not currently FDA-approved
1
OPTICAL QUALITY OF DUALOPTIC ACCOMMODATING
INTRAOCULAR LENS
Andrea Galvis, MD 1,3 a
Ivan Ossma, MD MPH MSc 1,2, a
Sanjeev Kasthurirangan, PhD 4, b
1 Fundacion Valle del Lili, Cali-Colombia
2 Universidad Industrial de Santander, Bucaramanga-Colombia
3 Universidad Autonoma de Bucaramanga, Bucaramanga-
Colombia
4 Abbott Medical Optics Inc, Irvine, USA
Financial Disclosure
a Consultant Abbott Medical Optics
b Scientist at Abbott Medical Optics Inc.
2
Synchrony Design
Anterior Optic 32
Diopters
Posterior Optic
Divergent (negative)
Study Purpose
+
+
Compare Optical Quality Pseudophakic Eyes
+
Monofocal IOL
+
Synchrony Dual Optic Accommodating IOL
Parameters
+
Objective iTrace aberrometry
+
+
Custom Matlab Software to calculate MTF
Subjective Contrast Sensitivity Testing
+
Optec 6500
3
Study Population
iTrace aberrometry
Cycloplegia
3 mm zone
Matlab Software to calculate MTF
Monofocal IOL : 15 subjects (30 eyes)
Synchrony IOL : 16 subjects (32 eyes)
4
5
iTrace - Matlab conversion
6
Results
Objective Measurements
3 mm pupil
Difference 0.01 um
7
Best-focus MTF
3 mm pupil
Differences are less than 10%
8
9
Mesopic Contrast Sensitivity
p=0.61 Wilcoxon Rank Sum Test
10
Discussion
Dual Optic Design DOES NOT affect optical or retinal image
quality of the eye
Objective evidence (iTrace, MTF)
Subjective evidence (CSF)
Contrast sensitivity of Synchrony is equivalent to that of
monofocal IOLs
Discussion
Effect of the slight spherical aberration on
Synchrony would only yield 0.42 D of increased
depth of focus
Calculations based on Rocha’s work *
The improved near vision of Synchrony patients
MOST LIKELY due to true accommodation
* Rocha KM. Using Adaptive Optics Simultion to Explore the Effect of Higher Order Aberrations on Depth of
Focus for the Correction of Presbyopia. Wavefront Congress 2008
11
12
Conclusion
Quality of Vision of Synchrony IOL similar to
standard monofocal IOLs
Measured Subjectively
Measured Objectively
Improved Near Vision with Synchrony
Not related to increased depth of focus