International Workshop on Burning Plasma Diagnostics
Varenna, Italy, September 24-28 2007
Design of a tomographic hard X-ray spectrometer for suprathermal electron studies with ECRH
S. Gnesin and S. Coda
École Polytechnique Fédérale de Lausanne (EPFL), Centre de Recherches en Physique des Plasmas (CRPP)
Association Euratom - Confédération Suisse, CH-1015 Lausanne, Switzerland
Introduction and motivations
The ECCD and ECRH system in TCV
TCV main parameters[4]:
Electron cyclotron resonance heating (ECRH) and current drive (ECCD)[1], disruptive
instability events and sawtooth activity have been demonstrated to produce suprathermal
electrons in fusion devices[2]. The importance of these phenomena for fusion reactors renders
suprathermal electron generation and dynamics a key topic in the physics of burning
plasmas[3]. Here some significant results from the TCV tokamak[4] are briefly reviewed and a
preliminary design of a novel tomographic hard X-ray spectrometer proposed for TCV is
discussed. The design is aided by simulations of tomographic reconstruction.
Overview of experimental results
X2 heating: 6 steerable launchers
X3 heating: 1 upper steerable launcher
Power: 0.5 MW each
Frequency: 82.7 GHz
Density limits: 4X1019 m-3
Pulse length: 2s
Power: 1.5 MW (3 gyrotrons)
Frequency: 118 GHz
Density limits: 1.1X1020 m-3
Pulse length: 2s
Proposed hard X-ray spectrometric system
Spectrometer: main parameters
RF field-particle resonance interaction
(ECRH; ECCD)
Suprathermal density
propagation in space
after short ECCD pulses
measured by HFS ECE
[7]
Suprathermal electron population is generated
Spatial resolution: 2 cm
Time resolution: down to 1 ms
Energy resolution: ~ 5keV at 60keV of photon
energy
Energy range: 20-200 keV
Detectors[10]: CdTe (or CZT)
Dimensions: 1mm X 1mm area X 2mm thickness
Detector per camera: ~ 35-40
Number of cameras: from 2 to 9
(trade-off between benefits for tomographic
reconstruction and costs)
Bremsstrahlung emission (hard X-rays)
due to electron-ion collisions.
ECE emission due to the Larmor motion
predominantly from suprathermals on HFS
C1
X2 generates suprathermal
electrons and contributes
to enhance the X3 power
absorption [5]
Local RF Power density achievable in
TCV > 20 MW/m^3
R =0.88m; a=0.23m; B0tor=1.45T; Ip<1MA;
k≤3; -0.7≤δ ≤0.7
C2
Compact and flexible camera design, wide
coverage for plasmas at different vertical
locations
Collimator design:
Divergent Soller collimator[11]
with radial detector array
C3
Detector
Line of sight
Fast electron broadening from
transport observed in many
ECRH discharges (resulting in
ECCD profile broadening) [6]
Tungsten metal foil enables good
photon stopping power with
compact size.
Upper spread border line
Lower spread border
line
C4
Collimator metal foil
C5
The HXR camera on loan from TORE SUPRA [8]
Now definitively reclaimed
Clearly evidenced the LFS-HFS asymmetry of
the poloidal bremsstrahlung distribution (possibly
related to trapped particles)
Other diagnostics used:
high-field-side electron
(ECE) radiometer [9]
oblique ECE
diamagnetic loop coil
cyclotron
C6
emission
C9
C8
C7
The design of the collimator aperture
can be adjusted for each single
detector in order to control the
étendue and its relative line spread in
the plasma region. The spatial
resolution of the system is related to
the typical chord separation in the
plasma region that is ~2cm. The
extreme compactness and flexibility of
this design enable it to be adapted to
other fusion devices.
New diagnostics being installed:
•Tangential HXR camera
•Vertical ECE
Tomographic reconstruction and diagnostic validation
Camera #1
2) Thin chord approximation
1) 3D Geometric integration
Tomography and poloidal mode number (m)
z
dΩ
L(ρ,φ)
Tomographic reconstruction enables to visualize
and determine quantitatively the detailed evolution
of the 2D shape of the emitting structure in the
plasma as a result of diffusion phenomena,
instabilities or other perturbations as well as of
ECRH and ECCD deposition.
5) Tomographic inversion and reconstruction
Chord brightness [a.u.]
Integrated
Measurements
Inversion
Methods
Slit aperture
(Aap)
dL
Δsx
m=2 doublet structure
(Three cameras)
4) Chord Signal integration
L
θd,p Collimator
θa,p
m=1 axial asymmetry
(Two cameras)
f i   g ( R, z )dl
Detector
Detector
(Ad)
dd,a
m=0 axial symmetry
(One camera)
y
3) Pixel Grid and T
matrix generation
R
Local emissivity
Li
Δstoroidal
g ( R, z)
Plasma
The power emitted by the
plasma along the chord
L(ρ,φ) is:
 Aap Ad cos a , p  cos d , p  
dP  
  g ( R, z )dL
2
4 d d ,a

 L
The tomographic algorithm that
generates the reconstructed plasma
emissivity uses the pixel grid method
and the minimum Fisher information
in order to condition the solution [12]
The chord f i  g  R, z dL
brightness:
Li
is independent of the particulars of
detector area, aperture size, etc. and
is therefore a convenient quantity to
express chord-integrated data in.

F  T1 * g
g Np  T NpNl  * FNl
i
Several tomographic reconstructions have been performed in order to validate the camera performance in recovering the shape and the intensity of simulated 2D plasma emission patterns. The simulations indicate as
expected an increasing in the quality of the tomographic reconstruction when cameras viewing the plasma from different directions are added.
Conclusions
Suprathermal electron dynamics is a
crucial element of ECRH physics and MHD
phenomena
important for fusion
Simulated
Emissivity
m=0
Cameras
2,5
Cameras
1,2,3,4,,5
Cameras
1,4,5
Simulated
Emissivity
m =2
Cameras
1,4,5
Cameras
1,2,3,4,5
A high-resolution, spectroscopic,
tomographic hard X-ray camera system is
being designed for the TCV tokamak
Simulations of tomographic
reconstruction from varying postulated
emissivity patterns underway to determine
the best detector distribution and to
optimize the reconstruction algorithm
References:
Simulated
Emissivity
m =1
Cameras
2,5
Simulated
Emissivity
(C)
Cameras
1,2,3,4,5
Cameras
1,4,5
Cameras
1,3,5
Cameras
1,2,4,5
Simulated
Emissivity
M=0 +m =2
+ Gauss peak
Cameras
1,4,5
Cameras
1,2,34,,5
[1].N. J. Fisch, Rev. Mod. Phys. 59, 175 (1987).
[2].P. V. Savrukin et al, Plasma Phys. Control. Fusion 48, B201 (2006).
[3].R. J. La Haye et al, Nucl. Fusion 46, 451 (2006).
[4].F. Hofmann et al, Plasma Phys. Conrol. Fusion 36, B277 (1994).
[5].S. Coda et al, 28th EPS conference on Control. Fusion and Plasma Phys.
Funchal, 18-22 June 2001. ECA Vol.25A, 301 (2001).
[6].S. Coda et al, Nucl. Fusion 43, 1361 (2003).
[7].S. Coda et al, Plasma Phys. Conrol. Fusion 48, B359 (2006).
[8].Y. Peysson, S. Coda, F. Imbeaux, Nucl. Inst. Meth. A 458, 269 (2001).
[9]. I. Klimanov et al, Plasma Phys. Control. Fusion 49, L1 (2007).
[10].T. Takahashi and S. Watanabe, IEEE Trans. Nucl. Sci. 48, 950 (2001).
[11].W. Soller, Phys. Rev. 24, 158 (1924).
[12].M. Anton et al, Plasma Phys. Control. Fusion. 38, 1849 (1996).