StatStrip - American Medical Technologists

Download Report

Transcript StatStrip - American Medical Technologists

Hospital Glucose Monitoring System

Hospital Glucose Meter Hospital Glucose Meter

Confidential

No. 260 StatStrip Glucose Rev 10/04/2012

Nova is the World Leader in Whole Blood Biosensors

Nova BioSensor Firsts 1978 First biosensor for Ionized Calcium 1980 First biosensor to measure Chloride in blood 1984 First biosensor to measure Total Calcium 1985 First biosensor to measure Hematocrit by ISE/Conductivity 1987 First biosensor to directly measure whole blood Glucose 1988 First biosensor to measure Lithium 1990 First biosensor to measure BUN (urea) 1992 First biosensor to directly measure whole blood Lactate 1994 First biosensor to measure Ionized Magnesium 1996 First biosensor to measure Creatinine 1996 First biosensor to measure Total CO 2 in whole blood 1997 First multi-wavelength fiber optic SO 2 % assay 1998 First fiber optic Hemoglobin assay 1998 First Ammonium biosensor 1999 First non-diluting direct Glutamine biosensor 1999 First non-diluting direct Glutamate biosensor 2001 First Acetate biosensor 2005 First Hematocrit corrected Glucose strip 2007 First calibration free Glucose strip 2007 First Creatinine strip and meter 2009 First calibration free Ketone strip 2009 First IgG assay on a biotechnology analyzer 2011 First Hematocrit corrected hospital Lactate meter and strip

Confidential

Nova Brand Products

Stat Profile ® pHOx Ultra Test Menu: pH, PCO 2 , PO 2 , SO 2 %, Hct, Hb, Na + , K + , Cl , Ca ++ , Mg ++ , TCO 2 , Glu, BUN, Creat, Lac, HHb, O 2 Hb, Methb, COHb, O 2 Ct, O 2 Cap, tBil Samples: Whole blood, Serum, Plasma Stat Profile ® pHOx Test Menu: pH, PCO 2 , PO 2 , SO 2 %, Hct, Hb, Na + , K + , Cl , Ca ++ , Glu, Lac Nova Chemistry Analyzers Test Menu: Na + , K + , Cl , TCO 2 , Ca ++ , Mg ++ , Li + , TCa, Glu, BUN, Crea, Hct, pH PO 2 Nova BioProfile FLEX , PCO 2 , Na + , K + , Ca ++ ™ Test Menu: Gluc, Lac, Gln, Glu, NH4+, pH, , CD, CV, Osm, IgG, PO 4 Samples: Whole blood, Serum, Plasma Samples: Whole blood, Serum, Plasma Samples: Cell Culture Media

*

Nova StatStrip ® Test Menu: Glucose Samples: Whole blood Nova StatStrip ® Test Menu: Glucose/Ketone Samples: Whole blood Nova StatStrip ® Test Menu: Lactate Samples: Whole blood Nova StatSensor ® Test Menu: Creatinine Samples: Whole blood Nova MAX ® Plus Test Menu: Glucose/Ketone Samples: Whole blood Nova MAX ® Link ® Test Menu: Glucose Samples: Whole blood Nova Lactate Plus Test Menu: Lactate Samples: Whole blood

* Currently Available outside the U.S.

®

Confidential

Hospital Laboratory Instrument Platform

Confidential

Hospital Laboratory Instrument Platform

Confidential

Hospital Laboratory Platform

Confidential

Hospital Point-of-Care Glucose Meter

Confidential

Hospital Point-of-Care Lactate Meter

Confidential

Hospital Point-of-Care Creatinine Meter

Confidential

Nova Max ® Plus Home Glucose Monitor Nova Max Link ® Home Glucose Wireless Monitor

®

Confidential

Professional Lactate Meter

Confidential

Nova Handheld Point-of-Care Hospital Analyzers

*

Confidential

TM

Nova Multi-Well Technology Creates A New Level of Analytical Performance

Hospital Glucose Meter Hospital Glucose Meter

Confidential

Hospital Glucose Meter The World’s Most Accurate Hospital Glucose Meter Proven in Over 60 Published Hospital Evaluations Worldwide

Laboratory quality accuracy and precision from 10 to 600 mg/dL Measures and corrects interferences from: - Hematocrit - Electrochemical interferences including acetaminophen, ascorbic acid, and uric acid No maltose, galactose, xylose or oxygen interference Eliminates calibration codes Fastest results, 6 seconds Small sample, 1.2 µL whole blood Nova Connects ™ , Total POC Connectivity Solution

StatStrip ® Hospital Connectivity Meter

Confidential

StatStrip

®

Has Earned Worldwide Scientific Acclaim

Confidential

StatStrip ® TM Multi-Well Technology Correlation Study N=1703 SLOPE: 1.018

5.6

11.1

INTERCEPT: -0.04 mmol/L INTERCEPT: -0.716 mg/dL (mmol/L) 16.7

22.2

27.8

R 2 : 0.9950

33.3

Sy*x = 0.54 mmol/L Sy*x = 9.795 mg/dL 700 600 500 400 38.9

38.9

33.3

27.8

22.2

300 200 100 0 0 100 16.7

11.1

5.6

200 300 (mg/dL) 400 Plasma Glucose Reference Method 500 600 700

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07

Confidential

N=559 300 200 100 0 0 700 600 500 400 Abbott PCx Correlation Study SLOPE: 0.89

5.6

11.1

INTERCEPT: 0.00 mmol/L INTERCEPT: 0.084 mg/dL (mmol/L) 16.7

22.2

R 2 : 0.9499

27.8

Sy*x = 1.84 mmol/L Sy*x = 33.194 mg/dL 33.3

38.9

33.3

27.8

22.2

100 200 300 (mg/dL) 400 Plasma Glucose Reference Method 500 600 16.7

11.1

5.6

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07

Confidential

N=225 600 500 400 300 200 100 0 0 Abbott PXP Correlation Study SLOPE: 0.820

5.6

11.1

INTERCEPT: 1.05 mmol/L INTERCEPT: 18.886 mg/dL (mmol/L) 16.7

22.2

R 2 : 0.97259

27.8

100 200 300 400 (mg/dL) Plasma Glucose Reference Method 500 Sy*x = 1.8 mmol/L Sy*x = 33.15 mg/dL 33.3

33.3

27.8

22.2

16.7

11.1

5.6

600

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07

Confidential

N=628 700 600 500 400 300 200 100 LifeScan SureStep Correlation Study SLOPE: 1.029

5.6

11.1

INTERCEPT: - 0.08 mmol/L INTERCEPT: - 1.505 mg/dL (mmol/L) 16.7

22.2

R 2 : 0.9869

27.8

33.3

Sy*x = 0.86 mmol/L Sy*x = 15.532 mg/dL 38.9

33.3

27.8

22.2

16.7

11.1

5.6

0 0 100 200 300 (mg/dL) 400 Plasma Glucose Reference Method 500 600

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07

Confidential

N=273 700 Roche Accu-Chek Correlation Study SLOPE: 1.036

5.6

11.1

INTERCEPT: 0.28 mmol/L INTERCEPT: 5.002 mg/dL (mmol/L) 16.7

22.2

R 2 : 0.9795

27.8

33.3

Sy*x = 1.18 mmol/L Sy*x = 21.197 mg/dL 38.9

600 33.3

500 400 300 27.8

22.2

16.7

200 100 0 0 100 11.1

5.6

200 300 (mg/dL) 400 Plasma Glucose Reference Method 500 600

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07

Confidential

StatStrip ® Correlation Study Comparison

Nova Abbott PCx Abbott PX I P Roche LifeScan N 1703 559 225 273 628

Summary

Slope 1.018

0.89

0.82

1.036

1.029

Intercept (mg/dL) -0.716

0.084

-1.505

R 2 0.995

0.9499

18.886

0.97259

5.002

0.9795

0.9869

Sy*x (mg/dL) 9.795

33.194

33.15

21.197

15.532

Kost, G. et al AACC, July 07; Diabetes Technology, Oct 07* * Summary table did not appear in publication

Confidential

StatStrip ® Multi-Well TM Technology Measures and Corrects Hematocrit Interference Hematocrit Interference Effect Hematocrit Interference Effect % Hct 25 Glucose Results Artificially High Up to 25% % Hct 45 (Calibration Point) Glucose Results Normal % Hct 65 Glucose Results Artificially Low Up to 30%

Confidential

StatStrip ® Multi-Well TM Technology Measures and Corrects Hematocrit Interference 30% 20% 10% 0% -10% -20% -30%

Error % Due to Uncorrected Hematocrit Glucose result uncorrected for Hct.

Glucose result.

Hct measured and corrected.

25% 45% Hematocrit 60%

Low hematocrit causes erroneously high glucose results High hematocrit causes erroneously low glucose results

Confidential

StatStrip ® Multi-Well TM Technology Measures and Corrects Hematocrit Interference Hematocrit Interference Study Nova StatStrip ® vs. Current Technology Test Strips

Roche Inform Abbott Precision Lifescan SureStep StatStrip 15 10 -5 5

0

-10 -15 -20 20 30

Glucose = 54 mg/dL

40 50

Hematocrit (%)

60 70

True value

Confidential

Karon, B. et al, Diabetes Technology and Therapeutics, Apr 2008

StatStrip ® Multi-Well TM Technology Measures and Corrects Hematocrit Interference Hematocrit Interference Study Nova StatStrip ® vs. Current Technology Test Strips

10

0

10 20 30 40 50 20 30

Glucose = 247 mg/dL True value

40 50

Hematocrit (%)

Confidential

60 70 Karon, B. et al, Diabetes Technology and Therapeutics, Apr 2008;

StatStrip ® Multi-Well TM Technology Measures and Corrects Hematocrit Interference Hematocrit Interference Study Nova StatStrip ® vs. Current Technology Test Strips

Roche Inform Abbott Precision Lifescan SureStep StatStrip 10

0

-10 -20 -30 -40 50 20 30 40 50

Hematocrit(%) Glucose = 486 mg/dL True value

60 70

Confidential

Karon, B. et al, Diabetes Technology and Therapeutics, Apr 2008

Severe hematocrit abnormalities are extremely common in hospitalized patients

 Up to 77% of ICU patients an have abnormally low hematocrit (<36%) during their ICU stay 1,2  Average HCT at ICU admission:   33-34% 2,3 29% have HCT<30% 4  HCT declines by an average of 1.5 - 2.0% HCT per ICU day 1 1. Von Ahsen N et. al., Important role of non-diagnostic blood loss and blunted erythropoietic response in the anemia of medical intensive care patients; Critical Care Medicine: 1999; 27(12); 2630-2639 2.Corwin HL et. al., The Crit Study: Anemia and blood transfusion in the critically ill; current practice in the United States;, Crit Care Med. 2004 Jan;32(1):30-52 3. T.S. Walsh et. al., Anemia during critical illness;, et. al., British Journal of Anaesthesia 2006, 97(3):278-291 4. Vincent JL, et al. Anemia and blood transfusion in critically ill patients. JAMA 2002 Sep 25; 288(12):1499-1507

Confidential

Maltose Interference with Glucose Dehydrogenase (PQQ Mediator) Enzymes

FDA issued two “Public Health Notices” regarding the dangers of GDH-PQQ glucose test strips.

Between 2004 and 2009,thirteen deaths have been reported directly related to the use of GDH PQQ glucose strips.

On August 13, 2009, the FDA clearly stated “

Avoid using GDH-PQQ glucose test strips in healthcare facilities

” http://www.fda.gov/MedicalDevices/Safety/Alert sandNotices/PublicHealthNotifications/ucm17699 2.htm

Confidential

344 280 217 Maltose Interference Studies Nova StatStrip ® vs Roche GDH-PQQ Method 290 Bed MA Hospital 150 Bed CT Hospital Maltose Interference

StatStrip

Maltose Interference

StatStrip GDH

180 163 120 108 54 54 0 100 Maltose (mg/dL) 200 Maltose Interference

StatStrip GDH

0 100 Maltose (mg/dL) Maltose Interference

StatStrip GDH

200 344 271 199 0 100 200 Maltose (mg/dL) Maltose Interference 0 100 Maltose (mg/dL) Maltose Interference

StatStrip GDH

200 540 460 380 522 450 380 0 100 Maltose (mg/dL) 200

Confidential

0 100 Maltose (mg/dL) 200

TM Nova StatStrip ® Multi-Well Technology Eliminates Calibration Codes

StatStrip lots are interchangeable Eliminates need for lot change over studies Reduces operator steps Eliminates glucose errors due to miscoding Glucose errors as large as 43% can be caused by an incorrect cal code

Baum JM et al. Diabet Technol & Ther 2006 Confidential

TM Nova StatStrip ® Multi-Well Technology Offers Other Operational Advantages

24 month shelf-life from manufacturer 180 day open vial stability Manufacturing lot size up to 8 million strips

Confidential

TM Nova StatStrip ® Multi-Well Technology Creates Significant Performance Improvements Top Layer

– Prevents biohazard exposure – Vent hole creates fast sample draw

Capillary Layer

– 1.2 µL sample fill – Design eliminates overfill errors

Measurement Wells Layer

Well 1: Interferences and glucose Well 2: Interferences Well 3: Hematocrit Well 4: Reference (and initiates analysis to eliminate short sampling errors)

Gold Layer

– Noble metal; superb electrochemical stability in all conditions – Classic electrochemical measuring surface – Excellent electrical properties Patent: US#6,287,451;6,837,976;EP#1 212 609;CA 2,375,092

Confidential

Glucose Management of Hospitalized Patients Demands Improved Glucose Strip Performance

JCAHO considers insulin to be one of the highest risk medicines in the inpatient setting “POCT glucose meters have now been employed for making therapeutically important insulin dosing decisions. It is essential that the accuracy of these measurements be comparable to those of laboratory analyzers..”

Khan A.I. 2006

“Careful assessment of glucose measurement and how it may impact the targets selected in the hospital are critical safety issues in intensive management of hyperglycemia.”

Dungan K. 2007

Khan AI et al. The Variability of Results Between Point-of-Care Testing Glucose Meters and the Central Laboratory Analyzer. Arch Pathol Lab Med, Vol 130, Oct 2006 Dungan K et al. Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management. Diabetes Care, Vol 30, Number 2, Februaly 2007

Confidential

Glucose Management of Hospitalized Patients Demands Improved Glucose Strip Performance

“The inaccuracies of glucose meters occur at the extremes of glucose reportable range (critical values), but in this region the most important and potentially life-threatening decisions concerning therapy must be made.”

Vasquez, J. 2005

“Published reports and clinical experience suggest that glucose meters may not be the most accurate and could possibly result in inappropriate insulin dose titration and avoidable adverse events related to glycemic control.”

Kanji, S. 2005

“The magnitude of the differences in the glucose values offered by the four different methods of glucose measurement led to frequent clinical disagreements regarding insulin dose titration in the context of an insulin infusion protocol for aggressive glucose control.”

Kanji, S. 2005

Kanji S et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med , Vol.33, No 12, 2005 Vasquez, J et al. Investigating the reliability of POCT glucose meters, Clinical Chemistry, Vol. 51, No. 6, Supplement, 2005

Confidential

Glucose Management of Hospitalized Patients Demands Improved Glucose Strip Performance

Glucose Error Leads To Insulin Dosing Error 1 Total Analytical Error Insulin Dosage Error Rate 5% 8-23% 10% 15% 16-45% >5% 2-Step Dosing Error 1. Boyd, JD and Bruns, DE. Quality Specifications for Glucose Meters: Assessment by Simulation Modeling of Errors in Insulin Dose.

Clinical Chemistry

, 47:2,209-214, 2001

Confidential

98% of Nova Results Satisfy a 10% Total Error Limit

N=133 patient study at Mayo Clinic Nova StatStrip, Lifescan SureStep, Abbott Precision, Roche Inform Versus Reference Method

StatStrip LifeScan Abbott Roche % Bias (Error)

≤

5% ADA

≤

10% ISO 15197

≤

15%

≤

20%

≤

25%

≤

30% Cumulative % of patients 75 98 100 Cumulative % of patients 35 71 94 100 26 56 76 93 100 Cumulative % of patients 35 67 80 96 98 100

Data analysis from Karon, B. et al, Diabetes Technology and Therapeutics, April 2008; not addressed in publication

ADA Specification ISO 15197 Requirement

Confidential

Fast and Easy Glucose Testing at the Bedside

Confidential

Strip Port Bright, Easy-to-Read Color Touchscreen. Large Easy-to-Use Touchpad. On/Off Button Bar Code Scanner

Confidential

Function Keys

StatStrip ® Connectivity Meter

Fastest Results, 6 seconds Sample Volume, 1.2 µL – Easier sample acquisition – Enables use of less painful lancets Fewest Operator Steps – No calibration codes or calibration chips – Strip lot entered automatically by default setting – Built-in 1D or 2D barcode scanner

Confidential

Color Touch Screen Prompts Operator Through Steps MED4 MED4

Confidential

Operator ID Entered by Scan or Touch Screen MED4

Confidential

Meter Then Goes Directly to Patient Test Screen MED4

Confidential

Strip Lot Scanned or Entered Automatically By Default Setting MED4 35475932

Confidential

Patient ID (or MRN) Entered by Scan or Touch Screen MED4 194546646

Confidential

StatStrip ® Confirms Valid Patient ID MED4 194546646

Confidential

Operator Prompted To Final Step MED4

Confidential

Results In Only 6 Seconds MED4

Confidential

Results screen displays all information – Patient ID – – – Strip Lot Time Operator ID Color, symbols and text allow easy recognition of normal, abnormal and critical ranges Operator can reject results at the meter – – Repeat testing Optional set-up

MED4

Confidential

Operator selects comments from a pull-down list or enters via free text Comments pull-down list customized by: – Meter location – QC, test or sample type Meters can be configured for mandatory comments Multiple comments can be entered for each result Comments can be designated as chartable or non-chartable

MED4 MED4

Confidential

Manual Test Entry

Enables electronic capture of non-connected POC tests using StatStrip infrastructure (ie urinalysis, pregnancy…) Undefined number of manual tests configurable Manual tests are configurable by hospital – pre-configured list for results – – Numeric, +/ Captures QC for manual tests Validate lot numbers

Confidential

StatStrip ® Docking Station Provides network connection for data downloading and meter configuration Charges batteries

Confidential

Comprehensive features for infection control

 StatStrip’s durable case has passed the FDA’s rigorous disinfection testing protocol validate with 10,950 cleaning cycles using commercially available cleaning materials including bleach.

 StatStrip’s disposable, protective bags are available for infection control. Patient barcode scanning and meter operation can be performed with the meter inside the protective bag.

 Software workflow and operating steps are optimized for infection control:  Operator ID and strip lot number are scanned and test strips pre-loaded into the meter outside the patient room  The carrying case and extra testing supplies remain outside the patient room per CMS recommendation.

Confidential

StatStrip ® Installation Examples Nova StatStrip

®

Healthcare Systems Include:

Intermountain Health Care System, UT (900+ meters) Manitoba Regional Healthcare, CN (855 meters) William Beaumont Medical Center, MI (379 meters) Johns Hopkins, MD (310 meters) Baptist Health South Florida, FL (250 meters) Northshore University Health System, IL (250 meters) Scripps Health, CA (400 meters) ROi, MO (1000 meters) Yale New Haven Health System, CT (500 meters) Duke University Health System, NC (400 meters) Kaiser Health, USA (2500 meters)

Confidential

StatStrip ® Installation Examples Largest Healthcare System Implementation

Intermountain Healthcare System, UT - 21 Hospitals - 60+ clinics - 900+ meters - 8,000+ operators trained

Confidential

Reimbursement

Confidential

StatStrip

®

CPT Code

Confidential

StatStrip

®

Reimbursement

StatStrip is the only glucose test strip developed specifically for professional use in the hospital, and cleared by the FDA for that intended use. The reimbursement code for StatStrip ® is 82947.

Two CPT codes for glucose: 82947 – Quantitative; glucose, blood $5.62 (StatStrip ® ) 82962 – Glucose; blood by monitoring device -$3.27

(FDA cleared specifically for home use) (All other glucose strips) Additional reimbursement for CPT Code #82947 $2.35

At an outpatient usage of 5%, hospital revenues would increase by the following amounts over a 5 year contract Five Year Strip Usage 2 million 4 million 6 million 8 million Approximate Meters in Use 80 160 240 320 Increased Total Hospital Revenue $235,000 $470,000 $705,000 $940,000

Confidential

Connectivity Overview

Confidential

Comprehensive, Cost Effective POC Connectivity Scalable connectivity from single hospital to multi-hospital systems Provides a single connectivity solution for all POC device types - Both wired and wireless connections - Manual test entry of off line tests Comprehensive data management and reporting tools Single device-type fee for StatStrip and StatSensor meters StatStrip and StatSensor meters eliminate the need for costly terminal server devices (Lantronix, Dawning).

Confidential

Enter Patient ID

Meter Name: NICU-1 1 4 7 Scan 2 5 8 0 Clear Op: 8636 13:42 3 6 9 ABC...

List

Meters Meter Configuration Data Management Lab/Hospital Information Systems

Nova Connects

Confidential

 Control of point-of-care devices, operators and usage for compliance with point-of-care program requirements  Capture of point-of-care data for analyzing, reporting and sharing  Connect point-of-care testing to multiple data receivers including LIS, HIS and EMR  Compatible with all “Open Connectivity” to all middleware providers

Confidential

Management and Control of Bedside Glucose Testing

NovaNet Instrument Manager StatStrip customized by location – ® meters can be Assign operators and privileges – Assign normal, abnormal and critical ranges – – Assign mandatory fields Establish download/docking requirements – Establish QC requirements • Pass/fail or numeric option, • QC frequency, lockout or prompting – Flag results for Tight Glycemic Control – – Reject results for repeat testing Structure patient ID or MRN

Confidential

Management and Control of Bedside Glucose Testing

NovaNet Instrument Manager Program coordinator can send a text message to specific operator or group of operators with an optional “I read this message” acknowledgment Multiple options for appending comments to patient or QC results – – – – By meter location By QC or test type Preset or free text Chartable or non-chartable

Confidential

NovaConnects: Manual Test Entry Manual test results can be entered through the StatStrip touch screen interface.

Data is captured through NovaConnects infrastructure Manual tests are configurable by the hospital - Manual tests can be quantitative or qualitative - QC is also captured An unlimited number of manual tests can be configured

Confidential

Integrated Management and Control of Bedside Testing Nova has worked with virtually all LIS and HIS vendors to provide fully integrated connectivity solutions that include: Complete data management and reporting Operator certification management Electronic (EDI) or scripted interfacing Full support for unsolicited result reporting to the LIS Hospital defined rules for result exception management Consolidated interface for all connected POC device types

Confidential

Capture all data for patient, device, operator and regulatory records Create standard or customized reports from a large report library Results exceptions Interface exceptions Results repository QA statistics Levy-Jennings Program utilization Operator Certifications Turnaround Time Results Reports Protocol Reports Program Summary Device Summary: Active devices, device location assignment, last upload Operations reports including;  Operator recertification due,  Daily workload,    Device error events, Overdue device docking, Reporting exceptions.

Confidential

Flexible network connections meet variable needs throughout hospital

Confidential

Nova Wireless Connectivity Options Wireless Bridge - Wireless connection from docking station provides access to Wifi network.

- Cost effective vs. installing new network drops - Uses Cisco Linksys WET-200, or other off-the-shelf components - Not Portable Nova Wireless Tote - Integrated docking station and transmitter in tote - “Always On” connection - Portable - No meter connectivity reconfiguration - Rechargeable battery

Confidential

Non-proprietary, Open Connectivity Enables connections to all POC devices, regardless of manufacture Eliminates connectivity lock-out which limits POC device choice.

Allows hospitals to make POC device decisions based on device needs and benefits Removes cost of middleware conversion as a barrier to acquiring new POC devices

Confidential

Nova Connects TM and Telcor, Inc. or Alere Informatics Proven success with virtually every LIS and HIS vendor Electronic Data Interface (EDI) or Script-in enables unsolicited results reporting into LIS or HIS Provides a single interface for virtually all POC devices from all vendors Manages data flow by hospital-specific exceptions rules

Confidential

Telcor QML

QML provides - Complete data management - ADT data messaging to NovaNet for positive patient ID at the meter - Hospital defined rules for result exception management - Consolidated interface for all connected POC device types

Confidential

Alere Informatics-RALS

RALS provides - Complete data management - ADT data messaging for positive patient ID at the meter - Hospital defined rules for results exception management - Consolidated interface for all connected POC device types

Confidential

TM StatStrip ® Multi-Well Technology Creates A New Level of Analytical Performance

“The StatStrip glucose meter

correlated best

with a plasma hexokinase reference method over a wide range of glucose concentrations and was

least significantly impacted by sample hematocrit and other interfering substances

. This should allow for

better management of critically ill patients

on intensive insulin therapy resulting in

better glycemic control

with improved outcomes and patient safety..”

Karon, B. et al, Diabetes Technology and Therapeutics, Apr 2008;

Confidential

Hospital Glucose Monitoring System

Hospital Glucose Meter Hospital Glucose Meter

Confidential Thank You