3200A
Options / Technical Overview
3200A Electrical Test Calibrator
Innovative design to bring advances in technology to the
calibration of Electrical Test Equipment
Calibrates Insulation Testers, RCD Testers, Loop Testers,
Multifunction Testers, PAT Testers, Flash & Breakdown
Testers
Transmille Ltd 1997 - 2011
3200A Feature Overview
Function
Comments
Continuity Resistance Source
HI Accuracy Option Available
Continuity Current Measurement
Insulation resistance Source
15kV Tester Option
Insulation test Voltage Measurement
16th and 17th Edition Compliant
AC Voltage Output
Continuously Monitored
RCD Test Time Measurement
RCD Test Current Measurement
Auto Waveform Analysis
RCD Test Duration Measurement
Loop Resistance Source
AutoLoop Option
PAT Earth Bond
Source Resistance & Measure Current to 30A
PAT Leakage
PAT Leakage Test Voltage Measurement
PAT Load Testing
Optional Adapter
PAY Flash Voltage Testing
Optional Adapter
PAT Flash Crrent Testing
Transmille Ltd 1997 - 2011
3200 Electrical Test
Equipment Calibrator –
Insulation Tester Calibration
Insulation Resistance from 100kohms to 10Gohms at 1kV
Measurement of Insulation Test Voltage at 1mA load
AC Voltage Calibration to 400 Volts
Continuity Resistance from 0 to 1kohms
Continuity Resistance Current to 300mA Measured into 1ohm
Calibrating Insulation testers with the 3200 is simple. All the functions required are in
one instrument, making calibration fast and simple.
Use with ProCal software to automate and save even more time.
3200 Electrical Test Equipment
Calibrator – RCD Testers
Residual circuit breaker testers are designed to test and measure the break
time of AC mains (RCD) trips at set test currents. Fault current is allowed to
pass between live phase and earth.
The ‘fault’ current is started by pressing a button on the tester, usually an LCD
counter on the tester measures the time in milliseconds until the mains supply
is disconnected.
The 3200A has been specially designed to make the
calibration of RCD testers safe & straight forward,
which can otherwise be quite difficult and even
dangerous.
To calibrate an RCD tester, it is necessary to measure the current taken from the mains supply by the
tester on each of its ranges and the time period measured shown on the testers display to the
disconnection of supply.
The 3200A can calibrate both functions. Intelligent firmware measures the correct fault current
ignoring pre-test currents from testers such as the AVO CM500. The 3200A also provides accurate
timed disconnection of the supply to verify the testers timing accuracy.
Some RCD testers have additional functions such as which half cycle of the mains the fault current
started, fault current duration etc. which can also be tested by the 3200A.
3200 Electrical Test
Equipment Calibrator –
Loop Tester Calibration
Automatic Measurement and compensation of
supply loop impedance
Automatic detection of faulty Tester preventing damage to 3200
8 loop values from 0.05ohm to 1kohm.
To calibrate a Loop Tester simple connect to the 3200, select the loop function and run
auto loop. Select the required loop value on the 3200 and press test. The 3200
automatically checks for a faulty loop tester before applying the full mains supply.
3200 Electrical Test
Equipment Calibrator –
PAT Calibration
Earth Bond Resistance
Earth Bond Current Measurement to 30 Amps
Insulation resistance from 1M to 10Gohms
Leakage Current Measurement
Class 1 & 2 Flash Voltage Measurement
There are several different measurement functions to be calibrated
on a PAT tester. The functions required are all in the 3200 ensuring
fast and efficient calibration of these instruments.
Increased Productivity
Reduced Test Times = Rapid Return On Investment
Transmille Ltd 1997 - 2011
External Options
External HiPot/Breakdown option
makes testing Breakdown testers
fast and safe.
EXTHV extends the resistance
range of the 3200 to calibrate
High Voltage insulation testers
safely and without the need for
expensive resistors
Transmille Ltd 1997 - 2011
3200 Electrical Test
Equipment Calibrator –
HIPOT / Breakdown Tester Calibration
3kV & 12kV AC/DC Measurement Ranges
200uA • 2mA • 20mA Current Ranges
200uA : 5M •10M •20M Loads
2mA : 500k •1M •2M Loads
20mA : 50k •100k •200k Loads
Current ranges : 1kV maximum
Power On & Overload Indicators
Requires 3200 Calibrator
VOLTAGE MEASUREMENT
Connects between ground and appropriate voltage terminal
Select voltage range (3kV / 12kV AC/DC) and read back the
voltage measured from the display of the 3200.
CURRENT LOAD
Three electronically selectable ranges, nine current
loads Select current range using the 3200 ‘softkeys’
Connect breakdown tester between ground and the
appropriate resistor socket.
3200A vs 5320A
• 3200 is Easier to use – Electrical socket on front
panel means direct connection for RCD and
Loop Testers
• Extended Workload – Using the external
options extend the workload of the 3200
• Ultimate Portability – 3200 weighs only 10Kg,
5320A weighs almost double
• Lower cost of initial investment
• External HiPot/Breakdown box simulates loads
as well as measures voltage
Transmille Ltd 1997 - 2011