pH and ICP Quality Control and Maintenance
Download
Report
Transcript pH and ICP Quality Control and Maintenance
pH and ICP Techniques
Clemson
Scoop 2 checks
One blank
Double scoop 6 samples
Tag set with lab
number, check, and
blank positions
pH quality control checks
1
2
3
Two QC check samples
Two samples at beginning, middle,
and end of set
Cross Check electrodes #1 and #2
Quality control row
ck
Check both QC check values at beginning and within set
Check all double scooped samples at beginning, middle,
and end of set
Buffer run - checks
Buffer run - blank (1:1, water:buffer)
End of run verification
Friday
◦ clean out electrode filling solution
◦ fill with water
Ultrasonic cleaner
Store in buffer pH 7 for the
weekend
Maintenance
Monday
◦ Take water out of electrode
◦ Refill with KCl filling solution
Verify flow of KCl through filter
Refresh buffer solutions
Electrode (from Fisher)
◦ Accumet rugged bulb pH
combination electrode
◦ 13-620-183A
Maintenance – cont.
5.3
5.25
5.2
5.15
5.1
5.05
5
0
20
40
60
pH Check values
80
100
120
140
Check B data since April 1, 2010
115 continuous reads over a 2.5 month period
Rinse pH < 4.5 and > 7.0
NAPT value 5.15
Lab average 5.15
stdev 0.036
Two checks within the set
One blank
ICP quality control checks
Sample entry
Probe – Leeman Labs
◦ Phone: 603-886-8400
◦ Part number 902-00017
Check all tubing for cleanliness and kinks
Change pump winding tube when needed
Adjust pump winding clamp - look for
smooth flow as sample travels though tube
Aspirate yttrium stock solution
Look for the tip of the red bullet
Adjust nebulizer gas flow rate
TJA
Nebulizer gas flow
ARCOS
ARCOS
Keep clean
◦ Take apart to rinse out nebulizer
and spray chamber parts
TJA – use cleaning wire
ARCOS
◦
SeaSpray Nebulizer
◦
Twister Spray Chamber
◦
Cross Flow
ARCOS
TJA
Nebulizer and spray chamber
ARCOS
ARCOS – clean as needed – salt buildup
TJA – clean ~once a year
Torch
TJA
25 seconds
P
K
Time (seconds) using pump speed of 60 rpm
Setting the flush time
ARCOS - Crossflow
Transient scan – run standard
Check to see when the peak intensities stabilize
Clemson – 35 second total flush
◦
25 seconds at 60 rpm
◦
10 seconds at 30 rpm
18 seconds
25 seconds
Crossflow spray chamber
Twister spray chamber
Transient scan on ARCOS
P – 100 ppm
Ca – 100 ppm
K – 100 ppm
Na – 50 ppm
25 seconds
45 seconds
Time in seconds using pump speed of 60 rpm
Transient scan going to rinse
ARCOS - Crossflow
Options for changing flush time
◦
◦
◦
◦
Spray chamber
Tubing length
Pump speed
Sample loop
Ca - 40 seconds
Pump speed 175 rpm
Run time scan
Check time for stability
Clemson soil method
◦ pump speed 200 rpm
◦ flush time 30 seconds
Setting the flush time - TJA
Ca – 396.847
Ca – 315.887
Saturated signal
Look at several recommended lines
Clemson - Ca line – 317.933 for soil
Line selection – ARCOS
Peak position
Set peak position
Select clean area
for background
subtraction
Look for
uncontaminated
blank
Cu – 324.754
Blank correction
blank
Method development - ARCOS
Run profile when needed
On the AROCS - check intensity of
P158.599 line
If significant UV intensity drop –
clean window and/or lens to optics
Check new standard against old
Read reference sample at start, mid
point, and end of run
Verify 2 check samples and blank
within set
Instrument profile and data
verification