What To Use When…..

Ken Kopp FCLSA Paragon Vision Sciences

Paragon CRT CRT Dual Axis RG-4

(Domestic only) All are laser marked

Paragon CRT

• Patented design using Proximity Control Technology • Certification necessary/up to -6.00 with up to -1.75 D of astig.

• Calculated from flat K and spherical manifest • 10.5 mm diameter is most often used/range from 10.0-12.0 mm • LZA has no radius of curvature • 100 lens DDS or SureFit options (domestic)

Paragon CRT Dual Axis

• This design mimics the shape of the astigmatic cornea by RZD and/or LZA manipulation without changing the BC (Proximity Control). • Based on the average elevation difference between the flat and steep meridian • FDA approval indication remains the same as the spherical product. Up to -6.00 D with or without up to -1.75 D of astigmatism • No certification necessary • Diagnostic sets of 16, 40 or 80 lenses

When Do I Use Which Design?

The CRT DDS should be the “go-to” tool for all reshaping candidates. It will be the design of choice to trial for proper centration, tear patterns, and correct BC selection.

When Do I Use Which Design?

• • • • • The pre treatment data is a “clue” as to what option choices you may need.

Flat K and spherical manifest Pre-treatment topography Refractive vs. Corneal astigmatism HVID Fissure and/or abnormal pupil size

Patient Pre-treatment Data

Flat K and Spherical Manifest • Flatter (<42.00 D) and steeper (>45.00) corneas can be more difficult to achieve proper centration and complete treatment.

Options – RG-4 or CRT • Higher myopes (>5.00 D) most often take longer to achieve stable VA and may need to treat in two steps.

Options – CRT or CRT Dual Axis depending on elevation differences

Patient Pre-Treatment Data

Topography Apical / Incomplete CRT or RG-4 Limbus-to-limbus Calculate elevation CRT Dual Axis

Patient Pre-Treatment Data

• • Refractive vs. Corneal Astigmatism More refractive than corneal = uncorrected astigmatism (ROL) More astigmatic correction than spherical component - not a desirable candidate • Corneal astigmatism approaching 2.00D will require a deeper sag (RZD/Mean K) choice initially. CRT DA/CRT/RG-4

Patient Pre-treatment Data

• HVID Average HVID is 11.8 mm • Larger diameters than 10.5 mm may be required to improve centration or aid patient comfort • Smaller diameters may be necessary for HVID of <10.8 mm CRT – 10.0 to 12.0mm, CRT DA – 10.5 & 11.0 mm, RG-4 = 10.0 to 12.4

Patient Pre-treatment Data

• Fissure and Pupil Size Asian fissure size and lid structure can determine diameter choice • • Larger pupil size (> 6mm ) in normal illumination can result in flare/glare issues Insertion/removal difficulties can determine design choice CRT/CRT DA/RG-4 have fixed 6 mm OZ

Design Choices

• CRT – Spherical With DDS, trial for all reshaping candidates .

• Make necessary parameter adjustments for proper positioning or edge lift appearance • Over-refract to determine the proper BC • If the desired position, “bull’s eye” pattern, edge lift and ROL are satisfactory – dispense and follow-up.

Design Choices

If the spherical CRT trial shows an incomplete “bulls eye” pattern or does not center properly…consider CRT Dual Axis Spherical CRT CRT Dual Axis

Incomplete/Apical Elevation Difference Spherical CRT Dual Axis CRT

CRT Dual Axis

Elevation difference • RZD difference depends on the calculated average elevation difference between the flat and steep meridian (pre-treatment) • RZD difference can range from 25 to 75 microns • LZA can also be altered in either meridian to adjust edge lift appearance – 2 degree maximum • If elevation data is not available, a default of 50 micron RZD difference with single LZA is chosen.

• Observe trial lens or order with calculated/default RZD

Design Choices

• Without DDS trial option Order CRT with *SureFit or normal warranty depending on pre-treatment criteria • If the Flat K is below 42.00 D or above 45.00 D and the spherical manifest is below -5.00 D, consider RG-4 first.

* SureFit has limited parameter specifications and only available in the US

Summary

The decision to choose a specific design is best derived from; • Spherical CRT trial lens • • • Proper position Proper “bull’s eye” pattern Over-refraction

Summary

• • If the CRT Trial lens does not display proper centration or incomplete “bull’s eye” pattern; Consider Dual Axis design based on elevation differences of the flat vs. steep meridian (from elevation/height map) Consider RG-4 for flatter corneas (<42.00 D) or steeper corneas (>45.00 D) or when the CRT design is no longer manageable.

Summary

Can either the CRT, Dual Axis , or the RG-4 lens be the initial lens choice and ordered empirically?

YES!!