Transcript Presentatie
30 januari 2014 Validatie bioanalyse volgens EMA-richtlijn Yvo de Beer Researchanalist MUMC+ 30 januari 2014 European Medicines Agency Opgericht in 1995, London Grotendeels betaald door farmaceutische industrie TAKEN Pre-autorisatie / Post-autorisatie geneesmiddelen Inspectie ( GMP / GLP ) Communicatie / Administratie ( documentatie ) 30 januari 2014 Traject opzet validatieprotocol Guideline on bioanalytical method validation Draft agreed by the Efficacy Working Party Adoption by CHMP for release for consultation End of consultation (deadline for comments) Agreed by Pharmacokinetics Working Party (PKWP) Adoption by CHMP Date for coming into effect September 2009 19 November 2009 31 May 2010 June 2011 21 July 2011 1 February 2012 30 januari 2014 Voorbeeld Comment Stakeholder 30 januari 2014 METHODEVALIDATIE HPLC GC MS VOLLEDIGE VALIDATIE PARTIELE VALIDATIE LIGAND BINDING IMMUNO ASSAY CROSS VALIDATIE VOLLEDIGE VALIDATIE PARTIELE en CROSS VALIDATIE VALIDATIERAPPORT ANALYSE MONSTERS HERANALYSE FOUTIEVE RUNS ANALYTISCH RAPPORT HERANALYSE AFGEHANDELDE RUNS 30 januari 2014 VOLLEDIGE VALIDATIE SELECTIVITEIT CARRY-OVER LLOQ CALIBRATIECURVE NAUWKEURIGHEID PRECISIE VERDUNNINGS EFFECT MATRIX EFFECT ( MS ) STABILITEIT 30 januari 2014 VOLLEDIGE VALIDATIE Uitvoeren in monstermatrix indien mogelijk. Gebruik van gecertificeerde referentiestoffen ( bv LCLABS, Lipomed, LGC-standards ). Analysecertificaat verplicht ( opslag condities, exp-datum, batch ) Interne standaard hoeft niet gecertificeerd te zijn ( geen anal. certif ) Bij MS isotoop gelabelde I.S. aanbevolen. Check voor unlabeling 30 januari 2014 VOLLEDIGE VALIDATIE SELECTIVITEIT Minstens 6 verschillende blanco matrices. Bij interferentie: response <20% LLOQ. <5% I.S. Co-medicatie Bij MS:conversie metabolietParent 30 januari 2014 VOLLEDIGE VALIDATIE CARRY OVER Injectie van zero sample na hoogste standaard tijdens validatie Concentratie <20 van LLOQ ( 5% I.S. ) Indien >20 % blanco’s injecteren 30 januari 2014 VOLLEDIGE VALIDATIE LOWER LIMIT OF QUANTIFICATION LLOQ = laagste standaard calibratiecurve Response LLOQ minimaal 5x response blanco LLOQ niet hoger dan 5% van Cmax 30 januari 2014 VOLLEDIGE VALIDATIE CALIBRATIE CURVE Range moet vooraf bekend zijn LLOQ=laagste standaard, ULOQ=hoogste standaard Minimaal 6 calibratiepunten 1 blanco ( zonder component(en) / interne standaard 1 “zero-sample”, met interne standaard Calibratiepunten mogen in duplo, maar is niet verplicht Rapporteer van curve: Slope, intercept, back-calculated concentraties Minimaal 3 curves rapporteren Calibratielijn hoeft niet vers gemaakt te worden. Calibratiepunten mogen opgeslagen worden ( bv bij -70°C ) en op een later tijdstip gebruikt worden. Stabiliteit moet in dit geval worden aangetoond 30 januari 2014 VOLLEDIGE VALIDATIE RAPPORTAGE CALIBRATIECURVE Rapporteer van curve: Slope, intercept, back-calculated concentraties Minimaal 3 curves rapporteren Back-calculated concentratie ijkpunten <15% Back calculated concentratie LLOQ <20% Minimaal 75% van calibratiepunten moeten voldoen aan criteria Bij duplo’s 50% per concentratieniveau moet voldoen Indien er niet wordt voldaan aan criteriaMethode herzien (aanpassen) 30 januari 2014 VOLLEDIGE VALIDATIE ACCURACY QC’s bereiden uit andere stockoplossing Minimaal 4 QC’s maken: LLOQ LOW QC (=< 3xLLOQ) MED QC ( +/- 50% calibratierange ) HIGH QC ( minimaal 75% van calibratierange ) Within run minimaal 5x opwerken per level Between run minimaal 3x opwerken per level Eisen Within run Gemiddelde berekende conc < 15% Gemiddelde berekenede conc LLOQ <20% Eisen Between run Berekende concentratie <15% Berekende concentratie LLOQ <20% Minimaal 1 run size-equivalent uitvoeren 30 januari 2014 VOLLEDIGE VALIDATIE PRECISION VC between run LLOQ <15% <20% VC within run <15% <20% LLOQ 30 januari 2014 VOLLEDIGE VALIDATIE DILUTION INTEGRITY Verdunning mag de nauwkeurigheid en precisie niet beinvloeden. De verdunningsintegriteit moet worden aangetoond door de matrix te spiken met analyt boven de ULOQ en deze te verdunnen met blanco matrix ( minstens 5x per verdunningsfactor ). De eisen die gelden voor de verdunde monsters zijn voor nauwkeurigheid en precisie beide 15% 30 januari 2014 VOLLEDIGE VALIDATIE MATRIX EFFECT ( Massaspectromterie ) Zeer uitgebreid verhaal ( paragraaf 4.1.8 ). I Hoofdpunten zijn: 6 verschillende blanco’s ( geen poolmonsters !! ) Bereken ratio spike in matrix / spike in oplossing ( analyt + I.S ) Uitvoeren bij lage conc. en hoge conc. ( max 3xLLOQ en bij ULOQ ) Wanneer bovenstaande niet uitvoerbaar is vanwege on-line sample prep. Check variatie in response. CV <15 30 januari 2014 VOLLEDIGE VALIDATIE STABILITY Stabiliteit aantonen met QCLOW en QC HIGH Direct bepalen en wegzetten onder opslagcondities QC’s bepalen op vers gemaakte ijklijn Berekende concentratie QC’s binnen 15% van nominale waarde Duur stabiliteit studie ≥ leeftijd monsters 30 januari 2014 VOLLEDIGE VALIDATIE STABILITY ( VERVOLG ) Stabiliteit stock en werk oplossingen ( analyten en interne standaard ) Vries/dooi stabiliteit analyt in matrixminimaal 12 uur in diepvries tussen cicli Korte termijn stabiliteit bij kamertemperatuur Lange termijn stabiliteit in vriezer ( -20 °C / -70 °C ) Indien mogelijk Stabiliteit opgewerkt monster bij kamertemperatuur (bv na droogdampen of in vial Stabiliteit in autoinjector 30 januari 2014 VOLLEDIGE VALIDATIE VALIDATIERAPPORT Opsomming gevalideerde parameters Beschrijving methode ( analyten, IS, monstervoorbwerking,extractie, analyse ) Referentiestandaarden Bron, batch nummers, COA, stabiliteit en opslag condities Run acceptatie criteria Tabel van alle runs met analysedatum ( goedgekeurd, afgekeurd, reden afkeuring ) Tabel van alle calibratieresultaten ( inclusief range, response functie, terugberekende concentraties en nauwkeurigheid Tabel van QC resultaten van alle geaccepteerde runs ( within en between run precisie en nauwkeurigheid. Waarden buiten de grenzen duidelijk markeren Stabiliteitsdata stock/werk oplossing en QC’s Selectiviteitsdata, LLOQ, Carry-over, matrix effect, verdunningsintegriteit Afwijkende resultaten en bijbehorende acties cq afwijking SOP 30 januari 2014 VALIDATION REPORT CAELYX® HIGH LEVEL VALIDATION REPORT CAELYX® HIGH LEVEL PROTOCOL No. UMCNONCO200902 PROTOCOL No. UMCNONCO200902 SUMMARY VALIDATION PERFORMANCES The following performances are tested: Selectivity (co-medication and interference endogenous components matrix), Stability autoinjector, Caelyx depegylation withinrun (dispersion test), Caelyx efficiency depegylation between-run (dispersion test), Between-run doxorubicine/doxorubicinol, Within-run doxorubicine/doxorubicinol, Freeze-thaw stability, 24 hour room temperature stability, Recovery, Stability longterm ( pending ) ASSAY PROCEDURE Doxorubicin, Doxorubicinol. Analytes: Daunorubicin Internal standard: REAGENTS / MOBILE PHASE Sodiumacetate-buffer:6,8 grams / liter water ( Na-Ac. 3 H2O ). pH=4.0 with Acetic acid Zinksulfate 89g/L: 89 grams of ZnS04.7H20 in liter water INSTRUMENTATION AND CONDITIONS Shimadzu LC-20AT HPLC-pump with DGU 20A3-degasser HPLC-system: Waters 717plus autosampler Jasco FP-920 Intelligent Fluorescence detector HPLC-column: Guard-column: Mobile phase: Needlewash: Flow: Excitation λ: Emmission λ: Bandwith: Gain: Temperature: Runtime: Injection volume: APPLIED ANALYTICAL METHOD Own method, based on articles Peter de Bruijn et al: Determination of doxorubicin en doxorubicinol in plasma of cancer patients by high-performance liquid chromatography ( Analytical Biochemistry 266, 216-221, 1999 ) AND David L. Chin et al: Rapid determination of PEGylated liposomal doxorubicin and its major metabolite in human plasma by ultraviolet-visible high-performance liquid chromatography ( Journal of Chromatography B, 779 259-269, 2002 ) Triton 6% (v/v): 6 ml Triton X-100 in 100 ml water. Heat to 40 °C to dissolve. 1. 250 μL Standard/Blank/QC/Sample + 25 μL Triton 6% (v/v) in eppendorf®cup. 2. Mix 2 minutes on vortex 3. Wait 10 minutes 4. Add 250 μL internal standard mix (IS + deproteinization solution). Except for blank 250 μL deproteinization solution without IS 5. Mix 2 minutes on vortex 6. Centrifuge 5 minutes at 10900 RPM 7. Transfer 250 μL in glas vial 8. Inject 5 μL onto HPLC-system Waters Atlantis T3, 15 x 4.6 mm; 5 μm Phenomenex Security Guard C18 4 x 3.0 mm Acetonitril : Sodiumacetate buffer 25:75 Acetonitril : Water 25:75 2,0 ml/min 480 nm 560 nm 40 nm 100 Ambient 22 minutes 5 μL Deproteinization solution: Zinksulfate 89g/L : Methanol ( 1: 1 ) Internal standard MIX: add 150 μL for every 3 mL of deproteinization solution REFERENCE STANDARDS COMPONENT ORIGIN BATCH NR CERTIFICATE STABILITY Doxorubicin Doxorubicinol Daunorubicin LC labs Alsachim Sigma DXR-102 CM-ALS-11-088-P3 1352629V YES YES NO 18-1-2016 21-9-2013 ? CALIBRATION STANDARDS AND QC SAMPLES Standard solution COMPONENT WEIGHT FACTOR C MATRIX PREPARATION PREPERATION DATE STORED Daunorubicin Doxorubicin Doxorubicinol .1 HCl .1 HCl .3 Fluoroacetate 1000 mg/L 1000 mg/L 1000 mg/L Methanol Methanol Methanol 9.77 mg in 10 mL 10.84 mg in 10 mL 12.55 mg in 10 mL 28-09-2012 28-09-2012 28-09-2012 5 °C 5 °C 5 °C Standard solution for QC samples COMPONENT WEIGHT FACTOR C MATRIX PREPARATION PREPERATION DATE STORED Doxorubicin Doxorubicinol .1 HCl .3 Fluoroacetate 1000 mg/L 1000 mg/L Methanol Methanol 10.80 mg in 10 mL 12.32 mg in 10 mL 05-11-2012 05-11-2012 NONE NONE VALIDATION REPORT CAELYX® HIGH LEVEL VALIDATION REPORT CAELYX® HIGH LEVEL PROTOCOL No. UMCNONCO200902 PROTOCOL No. UMCNONCO200902 Calibration results RUN ACCEPTANCE CRITERIA PARAMETER Selectivity: Co-medication Selectivity: Blank Matrix LLOQ Carry Over Back Calculated concentrations curve Within-run Accuracy Between-run Accuracy Within-run Precision Between-run Precision Stability autoinjector Stability Freeze-thaw Stability long therm Stability 24 hour STORAGE CONDITIONS -20 °C -20 °C 5 °C CRITERIA No interference Response <20% of response analytes LLOQ Response <5% of response IS Signal >5 times signal blank Response blank <20% of response analytes LLOQ Response blank <5% of response IS <15% for QC L,M,H <20 % for QC LLOQ <15% for QC L, M,H <20 % for QC LLOQ <15% for QC L, M,H <20 % for QC LLOQ VC <15% QC L,M,H VC <20% QC LLOQ VC <15% QC L,M,H VC <20% QC LLOQ <15% of nominal concentration <15% of nominal concentration <15% of nominal concentration <15% of nominal concentration PASSED YES YES Calibration date 15-11-2012 15-11-2012 12-11-2012 12-11-2012 13-11-2012 13-11-2012 14-11-2012 14-11-2012 REMARK List componentsExcel file YES YES Response 5.25 E-02 – 6.87 E-04 1.95 E-02 + 1.77 E-03 5.53 E-02 + 6.43 E-05 2.08 E-02 + 1.99 E-03 5.43 E-02 + 1.87 E-03 2.01 E-02 + 4.23 E-03 5.79 E-02 – 2.23 E-04 2.11 E-02 + 1.65 E-03 Back Cal. Concentration Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Accuracy Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet Excel sheet YES Concentrations in plasma ( derived from St 1000 mg/L ) YES CONC. DOXORUBICIN mg/L YES Blank Standard 0 Standard 1 Standard 2 Standard 3 Standard 4 Standard 5 Standard 6 QC 1 ( LLOQ ) QC 2 ( LOW ) QC 3 ( MED ) QC 4 ( HIGH ) YES YES YES YES PENDING YES ANALYSIS Analytical runs REMARK Range 0.5 – 50 mg/L DOXO 0.5 – 50 mg/L DOXOL 0.5 – 50 mg/L DOXO 0.5 – 50 mg/L DOXOL 0.5 – 50 mg/L DOXO 0.5 – 50 mg/L DOXOL 0.5 – 50 mg/L DOXO 0.5 – 50 mg/L DOXOL PARAMETER Selectivity: Co-medication Selectivity: Blank Matrix LLOQ Carry Over Back Calculated concentrations curve Within-run Accuracy Between-run Accuracy Within-run Precision Between-run Precision Stability autoinjector Stability Freeze-thaw Stability long therm ANALYSIS DATE 08-11-2012 09-11-2012 11/12/13/14-11-2012 14-11-2012 11/12/13/14-11-2012 PASSED YES YES YES YES YES 14-11-2012 11/12/13-11-2012 14-11-2012 11/12/13-11-2012 11-11-2012 14-11-2012 PENDING YES YES YES YES YES YES YES Stability 24 hour 14-11-2012 YES Depegylation within-run 14-11-2012 Depegylation between-run 15/16-11-2012 Recovery 11/12/13-11-2012 QC LLOQ QC LOW QC MED QC HIGH CONC. DOXORUBICINOL mg/L 0 0 0.5 1.5 5.0 10.0 25.0 50.0 0.5 1.5 25.0 40.0 0 0 0.5 1.5 5.0 10.0 25.0 50.0 0.5 1.5 25.0 40.0 DOXO WITHIN-RUN DOXO BETWEEN-RUN DOXOL WITHIN-RUN DOXOL BETWEEN-RUN 95.6% 92.6% 93.9% 96.6% 97.8% 93.3% 96.2% 99.3 % 87.6% 94.7% 97.3% 98.4% 92.2% 98.1% 99.6% 98.8% CV 2.9% CV 1.5% CV 2.3% CV 1.0% CV 1.2% CV 3.1% CV 1.3% CV 2.6% CV 3.4% CV 2.1% CV 1.6% CV 1.7% SUPPORTIVE DATA All not summarized data can be found in Excel-file: “Validatie hoge gebied Caelyx studie” YES YES AddedExcel file NO CRITERIA E.M.A VALIDATION PERFORMED ACCORDING TO: European Medicines Agency EMEA/CHMP/EWP/192217/2009 VALIDATION DONE BY: Y. de Beer REPORT DATE: 10-12-2012 AGREED : __________________ DATE: __________________ CV 8.7% CV 3.5% CV 0.6% CV 2.1% ANALYTISCHE RUN Minimaal 6 calibratiepunten in enkelvoud Blanco + zero sample Minimaal 3 QC’s in duplo ( onafhankelijk ) of minimaal 5% van het totaal aantal monsters Studie monsters in enkelvoud ( in duplo mag ) Acceptatiecriteria gelijk aan criteria validatie 67% QC’s moet voldoen (+minstens 1 van de duplo) ANALYTISCHE RUN Mogelijke redenen voor heranalyse Afwijking van criteria Duidelijk verschil IS bij monsters en standaarden Apparatuur probleem Afwijking in LLOQ of ULOQ waardoor range veranderd Analyt aanwezig in placebo Slechte chromatografie ANALYTISCHE RUN Heranalyse goedgekeurde runs Bij <1000 monsters10% heranalyseren Bij >1000 monster 5% heranalyseren Minder dan 20% afwijking t.o.v. gemiddelde voor 67% van alle monsters ANALYTISCH RAPPORT Vermelding referentiestandaard ( batch, COA, stabiliteit etc ) Opslag condities standaarden en QC’s Acceptatiecriteria Assay procedure Sample tracking ( codering monsters, opslag ) Tabel alle runs ( datum, calibratieresultaten, QC resulaten ) Afgekeurde runs ( datum, reden ) Andere afwijkingen Datasheet LINKS EMA WEBSITE Website: http://www.ema.europa.eu/ema/ Validatiedocument: http://www.ema.europa.eu/docs/en_GB/document_lib rary/Scientific_guideline/2011/08/WC500109686.pdf Commentaren op guideline: http://www.ema.europa.eu/docs/en_GB/document_librar y/Other/2011/08/WC500109687.pdf VRAGEN / DISCUSSIE