Current obstacles Between the Application of Ambient Ionization MS
Download
Report
Transcript Current obstacles Between the Application of Ambient Ionization MS
CURRENT OBSTACLES ON THE
APPLICATION OF AMBIENT
IONIZATION MS IN FORENSICS
Mike Rusak and Laura Fairburn
SUMMARY OF THE DRAWBACKS
• Cost
• Professional experience is needed
• Lack of comparable data
• Ionization is affected by spray, chemical, and surface
parameters.
• Problems with portability
SUMMARY OF THE DRAWBACKS
• Utilization of standards
• Problems with quantitative analysis
• Lack of research on matrix effects
• Lack of field testing
• Information obtained is not yet applicable to current
standards
COST
• Cost of the operator and the equipment can’t be
afforded by police departments
• The knowledge for just the calibration and
operation is very specialized, increasing one’s
expected salary
• Not a very practical purchase for a small police
department
• Right now, the prototypes of the portable MS
machines can only be afforded by the US military
EXPERIENCE NEEDED
• Need to have a lot of experience and education
• You can’t just walk off the street, be given
basic training and then be allowed to
operate the equipment and deemed an
expert
• Most evidence requiring MS is outsourced to
specialized forensic labs
LACK OF COMPARABLE DATA WHEN
WORKING WITH UNKNOWNS
• Since this technique is still in the prototype
phase, there aren’t many successful standard
cases to compare one’s results as few labs use
this technique
• All results need to be confirmed with other
techniques such as IR and NMR, or GC-MS
PARAMETERS THAT AFFECT IONIZATION
• All of the following parameters can be very difficult to
control in the field
1.Geometric parameters
2.Spray parameters
3.Chemical parameters
4.Surface parameters
PORTABILITY
• The large volume and power supply needed
reduces the portability of the machine
• Currently, portable machines are very rare and
experimental
• Compromises sensitivity and accuracy of
readings
UTILIZATION OF STANDARDS
• Standards are used to quantify and identify the analyte of
interest
• There’s a dependence of the ion signal on the type of
substrate present
• Eg) Thermal conductivity of substrate/matrix can
influence the signal
• Roughness of substrate
• Hard to quantify an analyte use a conventional MS
apparatus instead
UTILIZATION OF STANDARDS
• If it isn’t possible to mix in a standard, an intrinsic
compound of known concentration can be used, which
isn’t ideal.
• Not very quantitative
MATRIX EFFECTS
• Matrix: the material the analyte of interest is mixed with
• Influences desorption and ionization of the analyte
• Eg) cream, fillers (in the case of medicinal tablets),
dirt, carpet, etc.
• Systematic matrix studies are rare
MATRIX EFFECTS
• The matrix can cause overlap in the spectra due to
unwanted compounds
• Can also cause ion suppression
• Less volatile compounds in the matrix disrupt droplet
formation or evaporation, suppressing the amount of
ions reaching the detector
• Can react with analyte during desorption or ion
formation
MATRIX STUDIES
• Yu et al. studied the effect of different biological
matrices on the observed signal via DART analysis
• Showed that verapimil analyte signal can be reduced to
50% in bile, and 3% in brain tissue homogenate,
compared to verapimil in neat solvent
• Also showed that matrix effects are compound
dependent.
MATRIX STUDIES
• Jecklin et al. showed how the limit of detection (LOD)
varies when analyzing pesticides in different fruit juice
matrices
• The authors noted that background intensities were not
responsible for the observed changes
LACK OF FIELD TESTING
• Explosives, toxic industrial compounds, chemical
warfare agents, etc., have been explored by a variety of
methods
• Analysis of these compounds is limited for that fact that
the experiments are very controlled.
• Either done from standard solutions or complex
matrices doped with standards
LACK OF FIELD TESTING
• For drug analysis, pure chemicals are usually spiked
onto food, skin, or other surfaces.
• More research needs to be done with real evidence and
analytes in a non-standardized environment.
APPLICABILITY
• Yield a type of information which may not be applicable
to current standards
• Eg) Food policy requires that pesticide levels be
measured in concentration (wt/wt), which requires
homogenizing the fruit/veggies of concern.
• Sample preparation still needs to be done in some
cases
• Eg) To reduce ion suppression, especially for
analytes in a low concentration in complex matrices
REFERENCES
•
Alberici, Rosana M. et al. Ambient mass spectrometry: bringing MS into the “real world”.
Anal Bioanal Chem. 2010, 398: 265–294
•
Cooks et al., Forensic Applications of Ambient Mass Spectrometry. Anal Bioanal Chem
(2009) 394, 1995-2008
•
Chen et al., What Can We Learn From Ambient Ionization Techniques? J Am Soc Mass
Spectrom (2009) 20, 1947-1963
•
Gamez et al., Atmospheric pressure glow discharge desorption mass spectrometry for
rapid screening of pesticides in food. Rapid Commun Mass Spectrom. (2008) 22, 27912798
•
Takats, Zoltan et al. Ambient mass spectrometry using desorption electrospray ionization
(DESI): instrumentation, mechanisms and applications in forensics, chemistry, and
biology. J. Mass Spectrom. 2005, 40: 1261–1275
•
Yu et al., Bioanalysis without Sample Cleanup or Chromatography: The Evaluation and
Initial Implementation of Direct Analysis in Real Time Ionization Mass Spectrometry for
the Quantification of Drugs in Biological Matrixes. Anal Chem. (2009) 81,193-202