FieldSpec® HH UV/VNIR HandHeld Spectroradiometer
Analytical Spectral Devices, Inc. • • 5335 Stearling Drive, Suite A • Boulder, CO 80301 • • (303) 444-6522 fax:(303) 444-6825 www.asdi.com
Device Specifications • Range: 325-1075 nanometers • 512 channel silicon photodiode array overlaid with an order separation filter • Each channel will receive light in a 1.6 nanometer nominal band • Spectral resolution of approximately three nanometers at 700 nanometers
Optics • Light is collected through fiber optic head called a “bare head” • The bare head has a field of view (FOV) of 25 degrees. Foreoptics may be added to restrict this FOV.
• 10 and 1 degree FOV restricting foreoptics are included.
Pelican case is
not
waterproof
Thinking Inside the Box FOV 1 and 10 deg Shore Power Battery Serial to USB Cable Spectroradiometer “Help” Battery Charger Small White Reference
The Spectroradiometer
Serial to USB Cable
Charger RED Light: Charging GREEN Light: Charged FLASHING RED : Check connection
Mounting Device
Secure to Tripod • Remove quick release head from tripod • Insert only into bottom hole on device • Do not over tighten; only use fingers • Check tripod tight-points before trusting
Battery and Power Indicators Back of Device Power Battery Port Battery Port Shore Power Port Serial Port
FOV Device
FOV Limiting Foreoptics • Place threaded end on bare head • Do not over tighten • Use extreme care • Threads are located on end with label
Insert Battery and Turn On
Review • Don’t over tighten anything • 1.) connect cables and FOV limiting foreoptics • 2.) add battery and turn on by holding down
power
• 3.) now you are ready to attach to the computer!!
IMPORTANT • Do not plug the GPS into the serial port until you have turned on the computer. If the computer freezes you may need to turn off the ballpoint mouse in the device manager • Don’t let the spectroradiometer get too hot! If the spectra looks odd and it is hot out, first do a DC. If this will not help you will need to let it cool down.
Tough Book
Convert to Tablet (if needed)
Connections Tough Book Overview Controls Power “on”
On-Screen Keyboard Tough Book Controls Lock Rotate Display
Tough Book Connections Power Port USB for FieldSpec HH
Once FieldSpec HH is connected and on… For high-glare environments For normal lighting
White Reference and DC Buttons Raw DN and reflectance selector View Mode Buttons Status Bars Spectral View Area Save Status Bar GPS Display
Detail of Top Bar Pan mode allows for navigation while zoomed Zoom mode lets you examine spectra closer XY is normal view mode Freezes current spectra to allow examination Closes shutter
Set the Foreoptic First
Set the Integration Time
Integration Time • Adjust to improve signal strength • Avoid saturation (right), that will appear as off-graph and be displayed in red above sample status bar
CDA • Configurable Default Averaging will help to reduce noise • Higher number will take more time, but improve quality of signal • Control > Adjust Configuration
CDA Window 10 is default, but no less than 20 should be used.
• Always open Spectrum save before the first save • In this window set file name and other save settings • Control > Spectrum Save • Spacebar will save spectra
This information is encoded in every saved spectrum Amount / Frequency of save Save Location Save Name/Start Number Comments