Data Acquisition

Download Report

Transcript Data Acquisition

CT

Seeram Chapter 5: Data Acquisition in CT

Data Collection Basics

 X-ray source & detector must be in & stay in alignment  Beam moves (scans) around patient  many transmission measurements Patient X-Ray beams

Data Collection Basics

 Pre-patient beam  collimated to pass only through slice of interest  shaped by special bow tie filter for uniformity Filter Patient

Data Collection Basics (cont)

 Beam attenuated by patient  Transmitted photons detected by scanner  Detected photon intensity converted to electrical signal (analog)  Electrical signal converted to digital value  A to D converter  Digital value sent to reconstruction computer

CT “Ray”

 That part of beam falling onto a single detector Ray

Each CT Ray

 attenuated by patient  projected onto one detector  detector produces electrical signal  produces single data sample

CT View

 # of simultaneously collected rays

Scan Requires Many Data Samples  # Data Samples = [# data samples per view] X [# views]  # Data Samples = [# detectors] X [# data samples per detector]

Acquisition Geometries

 Pencil Beam  Fan Beam  Spiral  Multislice

Pencil Beam Geometry

 Tube-detector assembly translates left to right  Entire assembly rotates 1 o 1st Generation CT Tube 1 o Detector

Tube

Fan Beam Geometry

3nd Generation Detectors 2nd Generation 4th Generation

Comparing Long vs. Short Geometry Long Geometry • • • • • • • Smaller fan angle Longer source-detector distance Lower beam intensity Lower patient dose More image noise Less image blur Requires larger gantry Scan FOV Scan FOV

Spiral Geometry

    X-ray tube rotates continuously around patient Patient continuously transported through gantry No physical wiring between gantry & x-ray tube Requires “Slip Ring” technology Slip Rings Interconnect Wiring Tube Detector

What’s a Slip Ring?

Slip Rings  Electrical connections made by stationary brushes pressing against rotating circular conductor  Similar to electric motor / generator design

X-Ray Generator Configurations with Slip Ring Technology  Problem:  Supply high voltage to a continually rotating x-ray tube?

 Options  #1  Stationary Generator & Transformer  #2   Stationary Generator Transformer & x-ray tube rotate in gantry  #3  Transformer, generator & tube rotate in gantry

Option #1: Stationary High Voltage Transformer Incoming AC Power X-Ray Generator Primary Voltage Secondary Voltage High Voltage Transformer X-Ray Tube

Option #1: Stationary High Voltage Transformer Line Voltage Generator Primary Voltage HV Transformer  high voltage must pass through slip rings Secondary Voltage Tube Slip Rings Detector

Option #2: Rotating High Voltage Transformer Incoming AC Power X-Ray Generator Primary Voltage Secondary Voltage High Voltage Transformer X-Ray Tube

Option #2: Rotating High Voltage Transformer Line Voltage Generator Primary Voltage HV Transformer  low voltage must pass through slip rings Slip Rings Tube Detector

Rotating Generator

Incoming AC Power X-Ray Generator Primary Voltage Secondary Voltage High Voltage Transformer X-Ray Tube

Rotating Generator

 low line voltage must pass through slip rings Line Voltage Generator Slip Rings HV Transformer Tube

Spiral CT Advantages

 Faster scan times  minimal interscan delays  no need to stop / reverse direction of rotation  Slip rings solve problem of cabling to rotating equipment  Continuous acquisition protocols possible

X-Ray System Components

 X-Ray Generator  X-Ray Tube  Beam Filter  Collimators

X-Ray Generator

 3 phase originally used  Most vendors now use high frequency generators  relatively small   small enough to rotate with x-ray tube can fit inside gantry

X-Ray Tube

X-Ray Tube

 Must provide sufficient intensity of transmitted radiation to detectors  Radiation incident on detector depends upon  beam intensity from tube  patient attenuation   beam’s energy spectrum patient   thickness atomic #  density

Maximizing X-Ray Tube Heat Capacity     rotating anode  high rotational speed small target angle large anode diameter focal spot size appropriate to geometry  distances  detector size

Special Considerations for Slip Ring Scanners   continuous scanning means   Heat added to tube faster No cooling between slices Need  more heat capacity  faster cooling

Why not use a Radioactive Source instead of an X-Ray Tube?

   High intensity required  X-ray tubes produce higher intensities than sources Single energy spectrum desired   Produced by radioactive source X-ray tubes produce spectrum of energies Coping with x-ray tube energy spectrum  heavy beam filtering (see next slide)  reconstruction algorithm corrects for beam hardening

CT Beam Filtration  Hardens beam  preferentially removes low-energy radiation  Removes greater fraction of low energy photons than high energy photons   reduces patient exposure Attempts to produce uniform intensity & beam hardening across beam cross section Filter Patient

CT Beam Collimation

 Pre-collimators  between tube & patient Tube  Post-collimators • between patient & detector Detector

Pre-Collimation

 Constrains size of beam  Reduces production of scatter  May have several stages or sets of jaws Tube Pre-collimator Detector

Post-Collimation

 Reduces scatter radiation reaching detector  Helped define slice (beam) thickness for some scanners Tube Post-collimator Detector

CT Detector Technology: Desirable Characteristics  High efficiency  Quick response time  High dynamic range  Stability

CT Detector Efficiency

 Ability to absorb & convert x-ray photons to electrical signals

Efficiency Components

 Capture efficiency  fraction of beam incident on active detector  Absorption efficiency  fraction of photons incident on the detector which are absorbed  Conversion efficiency  fraction of absorbed energy which produce signal

Overall Detector Efficiency

Overall detector efficiency = capture efficiency X absorption efficiency X conversion efficiency

Capture Efficiency

 Fraction of beam incident on active detector

Absorption Efficiency

Fraction of photons incident on the detector which are absorbed  Depends upon detector’s   atomic # density   size thickness  Depends on beam spectrum capture efficiency X absorption efficiency X conversion efficiency

Conversion Efficiency

 Ability to convert x-ray energy to light GE “Gemstone Detector” made of garnet

Conversion Efficiency

 Ability to convert x-ray energy to light Siemens UltraFastCeramic (UFC) CT Detector • Proprietary • Fast afterglow decay UFC Material UFC Plate

Response Time

 Minimum time after detection of 1st event until detector can detect 2nd event  If time between events < response time, 2 nd detected event may not be  Shorter response time better

Stability

 Consistency of detector signal over time  Short term  Long term  The less stable, the more frequently calibration required

Dynamic Range

 Ratio of largest to smallest signal which can be faithfully detected  Ability to faithfully detect large range of intensities  Typical dynamic range: 1,000,000:1  much better than film

Detector Types: Gas Ionization

 X-rays converted directly to electrical signal Filled with Air

X-Rays +

Ionization Chamber

-

- + Electrical Signal

CT Ionization Detectors

 Many detectors (chambers) used  adjacent walls shared between chambers  Techniques to increase efficiency  Increase chamber thickness  x-rays encounter longer path length  Pressurize air (xenon)  more gas molecules encountered per unit path length

X-Rays

thickness

Older Style Scintillation Detectors

 X-rays fall on crystal material  Crystal glows  Light flash directed toward photomultiplier (PM) tube  Light directed through light pipe or conduit  PM tube converts light to electrical signal  signal proportional to light intensity PM Electrical Signal

Detector Types: Scintillation

 X-ray energy converted to light  Light converted to electrical signal

X-Rays

Light Scintillation Crystal Photomultiplier Tube Electrical Signal

Photomultiplier Tubes

 Light incident on Photocathode of PM tube  Photocathode releases electrons + -

X-Rays

Scintillation Crystal Light Photocathode Dynodes PM Tube

Photomultiplier Tubes

 Electrons attracted to series of dynodes  each dynode slightly more positive than last one + + + +

X-Rays

Scintillation Crystal Light Photocathode + Dynodes PM Tube

Solid State Detectors

 Crystal converts incident x-rays to light  Photodiode semiconductor current proportional to light

X-Rays

Light Photodiode Semiconductor Electrical Signal

Photodiode

 Made of two types of materials  p-type  n-type  Lens focuses light from crystal onto junction of p & n type materials

X-Rays

Light Lens p n Junction

Photodiode

 Light controls resistance of junction  Semiconductor current proportional to light falling on junction

X-Rays

Light Lens p n Junction

Solid State Detectors

 Output electrical signal amplified  Fast response time  Large dynamic range  Almost 100% conversion & photon capture efficiency  Scintillation materials   cadmium tungstate high-purity ceramic material

Detector Electronics

From Detector Pre-Amplifier Increases signal strength for later processing Logarithmic Amplifier Analog to Digital Converter To Computer Compresses dynamic range; Converts transmission intensity into attenuation data

Logarithms

Log 10 x = ? means 10 ?

= x?

 logarithms are exponents  log 10 x is exponent to which 10 is raised to get x  log 10 100 =2 because 10 2 =100

Logarithms

Input 100,000 10,000 1,000 100 10 1 Logarithm 5 4 3 2 1 0 Using logarithms the difference between 10,000 and 100,000 is the same as the difference between 10 and 100

Compression

1,000 3 = log 1000 2 =log 100 1 = log 10 0 = log 10 Hard to distinguish between 1 & 10 here 1 10 100 1000 Input 100,000 10,000 1,000 100 10 1 Logarithm 5 4 3 2 1 0 1 10 100 1000 Difference between 1 & 10 the same as between 100 & 1000 Logarithms stretch low end of scale; compress high end

Logarithmic Amplifier

 accepts widely varying input  takes logarithm of input  amplifies logarithm  logarithm output dynamic range now appropriate for A/D conversion Input Logarithm 100,000 10,000 1,000 100 10 1 5 4 3 2 1 0

Improving Quality & Detection

 Geometry   Smaller detectors Smaller focal spot    Thinner slices  Larger focus-detector distance Smaller patient-detector distance less patient variation over slice thickness distance