Signal to noise ratio Contrast to noise ratio Spatial resolution Scan time Volume imaging
Introduction
The choice of pulse sequence determines the weighting and the quality of the images as well as their sensitivity to pathology The timing parameters selected determines the weighting of the images The quality of the image is controlled by – Signal to noise ratio (SNR) – Contrast to noise ratio (CNR) – Spatial resolution – Scan time
Signal to Noise Ratio (SNR)
Signal to noise ratio (SNR)
SNR = amplitude of the signal / average amplitude of the noise The signal is the voltage induced in the receiver coil by the precession of the NMV The noise is generated by the presence of the patient in the magnet, and the background electrical noise of the system
Noise
Is constant for the patient Depends on the – build of the patient – Area under examination – Inherent noise of the system Occurs at every frequency and is random in time
Signal
Is cumulative and depends on many factors and can be altered Increasing the signal decreases the SNR Any factor that affects the signal amplitude in turn affects the SNR
Factors that affect SNR
Proton density of the area under examination Voxel volume TR, TE and flip angle NEX (number of excitations) Receive bandwidth Coil type
Proton density
Areas with low proton density (PD) produce low signal and therefore low SNR – E.g. Lungs, cortical bone Areas with high PD have high signal & high SNR – E.g. Bladder, renal pelvis
Voxel volume
Large voxels have more nuclei therefore high signal and high SNR Small voxels have less nuclei therefore low signal and low SNR Any parameter that change the voxel volume changes the SNR – Voxel volume = pixel area x slice thickness – Pixel area = FOV dimensions/matrix size
Voxel Volume
voxel Small voxel few spins Matrix slice Slice width Large voxel more spins & high signal
TR, TE and flip angle
Spin echo pulse sequences have more signal than gradient echo sequences, because; – All the longitudinal magnetization is converted into transverse magnetization by 90 0 flip angle – Gradient echo pulse sequences only convert a portion of the longitudinal magnetization into transverse magnetization as the flip angle is not 90 0 – The 180 0 rephasing pulse is more efficient at rephasing than the rephasing gradient of gradient echo sequences
The lower the flip angle , the lower the SNR Long TR increases SNR and a short TR reduces SNR (TR controls the amount of longitudinal magnetization that is allowed to recover) Long TE reduces SNR and short TE increases SNR (TE controls the amount of transverse magnetization that is allowed to decay before an echo is collected)
Number of signal averages (NSA) or NEX
This is the number of times data is collected with the same amplitude of phase encoding slope.
Increase in NEX increase the amount of data stored in each line of k-space.
The data consists of both signal and noise.
The noise increases randomly and the increase of signal is not random.
Therefore the increase of NEX increases the SNR by √2(=1.4)
NEX Vs SNR
SNR 3 2 1 1 2 3 4 5 6 7 8 NEX
Receive bandwidth
This is the range of frequencies that are sampled during the application of readout gradient.
Reducing the receive bandwidth results in less noise being sampled relative to signal.
Therefore SNR increases bandwidth is decreased.
when receive But Sampling tine increases minimum TE available and increases the Increases the chemical shift artefact
Bandwidth versus SNR
noise
Bandwidth +/- 16 kHz
signal
noise noise noise Bandwidth +/- 4 kHz
signal
noise noise SNR increases with decrease of bandwidth
Type of coil
The type of coil used affects the amount of signal received and therefore the SNR Volume coils with quadrature excitation two coils are used to receive signal increase SNR as Surface coils placed close to the area under examination also increase the SNR The use of the appropriate receiver coil plays extremely important role in optimizing SNR The volume of tissue sensitive volume imaged should optimally fill of the coil the Large coils increase the likelihood of aliasing outside the FOV more likely to produce signal as tissue
RF coil types
Commonly used – Volume coil or Bird-cage coil – Surface coil – Phased array coil Specific – Solenoidal coil – Helmholtz pair
How to increase SNR
Use spin echo sequences where possible Try not to use a very short TR and a very long TE Use the correct coil and ensure that it is well tuned Use a coarse matrix Use a large FOV Select thick slices Use as many NEX as possible
FOV, Resolution & SNR
FOV = 24cm High SNR FOV = 12 cm High resolution, low SNR
Slice thickness, Resolution & SNR
10 MM SLICE 3 MM SLICE High resolution, low SNR
NEX & SNR
NEX = 4, TIME= 6 MIN High SNR PARTIAL AVERAGING, TIME 56 S
Contrast to noise ratio (CNR)
CNR
Defined as the difference in the SNR between two adjacent areas Controlled by the same factors that affect SNR Most critical factor affecting image quality Determines the eyes ability to distinguish areas of high signal from areas of low signal (Although the SNR of T2 weighted image is lower than T1 weighted image the ability to distinguish tumour from normal tissue is greater because of the high signal compared to the low signal of surrounding anatomy, i.e. CNR is higher)
Image contrast depends on: – TR – TE – TI – Flip angle – Flow – Turbo factor (in fast spin echo) – T1 – T2 – Proton density – Magnetization transfer coherence (MTC)* * (see note)
Spatial Resolution
This is the ability to distinguish between two points as separate and distinct.
Controlled by the voxel size Small voxels results in good spatial resolution The voxel size is affected by – Slice thickness – FOV – Number of pixels or matrix The spatial resolution is increased by using rectangular FOV for rectangular anatomy Rectangular FOV reduces the scan time
Matrix & resolution
Matrix 256 x 256 Matrix 512 x 256
Scan time
Short scan times are important to minimize the possibility of patient movement Factors that affect the scan time are – TR – time of each repetition – Number of phase encodings – number of lines of K space – NEX – number of times data is collected with the same phase encoding gradient
How to reduce the scan time
Use the shortest TR possible Select the coarsest matrix possible Reduce the NEX to a minimum
Trade-offs
There are many trade-offs when selecting parameters within a pulse sequence Ideally an image should have – high SNR, – Good spatial resolution – a very short scan time.
But when improving one factor inevitably reduces one or both of other two
Decision making
Selections depend on – The area to be examined – Condition and co-operation of the patient – Clinical throughput required
Tips to improve image quality
Choose the correct coil Make sure that the patient is comfortable ad immobilized Ascertain from the radiologist what sequences are required SNR is the most important quality factor Keep the scan time as short as possible
Volume Imaging
Advantages to demonstrate very small lesions Slice thickness can be reduced Entire volume of tissue is excited & No slice gap SNR is superior Fewer NEX can be used Can look at anatomy in any plane
The scan times are relatively long therefore used in conjunction with faster pulse equences Slices are sectioned out by a series of phase encoding steps along the slice select axis Therefore the scan time increases But as SNR is increased with increased number of slices the NEX can be reduced Isotropic (equal dimensions) pixel give equal resolution in any plane Scan time = TR x NEX x number of phase encodings x number of slice encodings
Conclusion
Manipulating SNR, image contrast, spatial resolution and scan time is a real art and takes some time and experience Even after many years you may get things wrong occasionally Perseverance is important to get good image quality